2-Wavelength Fluorometer

Reference Designator
GP05MOAS-GL365-01-FLORDM000
Start Depth
0
End Depth
1,000
Location
Current Status
Review Complete
Class
FLORD (2-Wavelength Fluorometer)
Series
FLORD-M
Science Discipline
Biological
Make
WET Labs
Model
ECO Puck FLBB-SLC
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/GP05MOAS/GL365/01-FLORDM000/metadata
Method Data Stream Content Type
recovered_host flord_m_glider_instrument_recovered Data Products Report M2M Stats Science
telemetered flord_m_glider_instrument Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review MV-1309 07/19/2013 02/21/2014 CGVGP-05MOAS-00000 CGVEH-GLDROO-00365 CGINS-FLORDM-03124 50.0866 -144.449 1000 1000
2 Review RB-16-05 06/27/2016 05/22/2017 CGVGP-05MOAS-00000 CGVEH-GLDROO-00365 CGINS-FLORDM-03124 50.1656 -144.501 1000 1000
Metadata Start Date End Date Comment
GP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

GP05MOAS-GL365
Deployment: 1

latitude and longitude are missing from all deployment 1 .nc files (Redmine #14657)

By Lori Garzio, on 3/16/20

GP05MOAS-GL365
Deployment: 1

Based on the data, this glider looks like it had operational issues during the second half of the deployment 1. This should be annotated.

By Lori Garzio, on 3/17/20

GP05MOAS-GL365-01-FLORDM000
Deployment: 1

43% of deployment 1 optical_backscatter values are NaNs

By Lori Garzio, on 3/17/20

New Note

Metadata Start Date End Date Comment
GP05MOAS-GL365
6/27/16, 11:35 AM 5/1/17, 6:53 AM

Deployment 2: Glider recovered by a ship of opportunity (R/V Sikuliaq) due to limited battery power. Science sensors were turned off on 1/27/2017 to conserve battery, then resumed sampling around 4/2/2017 to 5/1/2017.

Id: 594 By: lgarzio