3-Wavelength Fluorometer

Reference Designator
GS05MOAS-PG565-03-FLORTM000
Start Depth
0
End Depth
200
Location
Current Status
Review Complete
Class
FLORT (3-Wavelength Fluorometer)
Series
FLORT-M
Science Discipline
Biological
Make
WET Labs
Model
ECO Puck FLBBCD-SLK
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/GS05MOAS/PG565/03-FLORTM000/metadata
Method Data Stream Content Type
recovered_host flort_m_sample Data Products Report M2M Stats Science
telemetered flort_m_sample Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review NBP-15-11 12/20/2015 11/21/2016 CGVGS-05MOAS-00000 CGVEH-GLDRGP-00565 CGINS-FLORTM-03815 -54.1252 -89.7625 200 200
Metadata Start Date End Date Comment
GS05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

GS05MOAS-PG565-03-FLORTM000

Comparison with the shipboard CTD cast was done for D0001. Chl-a recorded by the glider was generally comparable to the CTD cast, with the exception of the chl-a maximum (~25 - 50 dbar) and at deeper depths (>100 dbar) where the glider recorded higher chl-a values. Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/11/19

GS05MOAS-PG565-03-FLORTM000

Biofouling could be an issue in the second half of the deployment.

By Lori Garzio, on 1/29/20

New Note

Metadata Start Date End Date Comment
GS05MOAS-PG565
12/20/15, 12:56 PM 11/5/16, 9:02 PM

Deployment 1: Deployed on 12/20/2015, recovered on 12/25/2015 20:40 UTC for ballast correction and was re-deployed. Starting around 5/1/2016, glider became increasingly difficult to steer. Glider stopped profiling on 11/6/2016 to conserve remaining battery power. At recovery, barnacles were present mostly on the underside of the glider, and were not in an area to hinder flight performance.

Id: 577 By: lgarzio