    GI05MOAS-GL484-02-DOSTAM000

    Comparisons with the shipboard CTD casts were done for D0002 and D0003. The shipboard CTD cast data are corrected for temperature, salinity, and depth (Redmine 14230). The L2 DO data product (corrected for temperature, salinity and depth) is substantially lower for the glider compared to the shipboard data. This could be related to the issue of incorrect default salinity settings on some glider optodes (Redmine 13575). Comparison to the shipboard bottle data is outside the scope of this project. Note: the conversion from ml/l to umol/kg for the shipboard DO = DO (ml/l) * 44.661 / 1.025.

    By Lori Garzio, on 3/8/19

    GI01SUMO-RID16-02-FLORTD000
    Deployment: 3
    8/12/17, 9:00 PM 8/13/17, 2:00 PM

    There are 2 days at the end of deployment 3 for which data are available in the telemetered stream and not the recovered_host stream. These data should be available in the recovered_host stream.

    By Lori Garzio, on 3/22/19

    GI01SUMO-RID16-02-FLORTD000
    Deployment: 2
    2/8/16, 12:00 AM 7/19/16, 12:00 AM

    Data look suspect shortly after annotation ID 47 indicates violent weather caused multiple power outages. This time period should be investigated and annotated. Excluding from data range summary.

    By Lori Garzio, on 3/22/19

    GI01SUMO-RID16-02-FLORTD000

    The automated valid data tests were triggered for all deployments because of the high percentage of fill values for seawater_scattering_coefficient and optical_backscatter.

    By Lori Garzio, on 3/22/19

    GI01SUMO-RID16-02-FLORTD000

    pressure_depth (dbar) is an array of fill values in all files. This variable should contain valid pressure data from the collocated CTD.

    By Lori Garzio, on 3/22/19

    GI01SUMO-SBD12-02-FLORTD000

    pressure_depth (dbar) is an array of fill values in all files. It's unclear if this variable should contain valid data, or if it's assumed that the pressure is 1 because this instrument is on the surface buoy.

    By Lori Garzio, on 3/21/19

    GI01SUMO-RID16-02-FLORTD000
    Deployment: 1
    9/10/14, 6:50 PM 8/22/15, 11:00 AM

    Annotation ID 1521 indicates that the instrument flooded and there is no valid data from this deployment. Excluding from data range summary.

    By Lori Garzio, on 3/22/19

    CP02PMCO-WFP01-03-CTDPFK000
    Deployment: 1
    6/28/14, 12:00 AM 7/22/14, 12:00 AM

    suspect data profile

    By Leila Belabbassi, on 3/21/19

    CP02PMCO-WFP01
    12/6/14, 12:00 AM 5/2/15, 12:00 AM

    bad data for CTD, DO, FLORT, VEL3D and PAR seen after 15 December 2014 to the end of the deployment.

    example plot from CTD shown.

    By Leila Belabbassi, on 3/21/19

    CP02PMCO-WFP01
    12/6/14, 12:00 AM 5/2/15, 12:00 AM

    annotation ID 712 should be updated by adding that data quality is bad for this time range.

    By Leila Belabbassi, on 3/20/19

    GI01SUMO-SBD12-02-FLORTD000
    Deployment: 3
    8/12/17, 9:00 PM 8/14/17, 4:00 AM

    There are 3 days at the end of deployment 3 for which data are available in the telemetered stream and not the recovered_host stream. These data should be available in the recovered_host stream.

    By Lori Garzio, on 3/21/19

    GI01SUMO-SBD12-02-FLORTD000

    The automated valid data tests were triggered for all deployments because of the high percentage of data outside of global ranges and outliers (for all variables).

    By Lori Garzio, on 3/21/19

    GI01SUMO-RII11-02-FLORDG033
    Deployment: 4

    Total volume scattering coefficient is substantially higher for deployment 4 compared to the previous 2 deployments. A similar pattern is observed on the other two fluorometers on RII11 (FLORDG031 and FLORDG032). This should be investigated.

    By Lori Garzio, on 3/21/19

    GI01SUMO-RII11-02-FLORDG032
    Deployment: 3
    10/10/16, 12:00 AM 11/27/16, 4:04 PM

    Instrument appeared to start failing (negative total volume scattering coefficient values). Excluding from data range summary.

    By Lori Garzio, on 3/21/19

    GI01SUMO-RII11-02-FLORDG032
    Deployment: 4

    Total volume scattering coefficient is substantially higher for deployment 4 compared to the previous 2 deployments. A similar pattern is observed on the other two fluorometers on RII11 (FLORDG031 and FLORDG033). This should be investigated.

    By Lori Garzio, on 3/21/19

    GI01SUMO-RII11-02-FLORDG031
    Deployment: 4

    Total volume scattering coefficient is substantially higher for deployment 4 compared to the previous 2 deployments. A similar pattern is observed on the other two fluorometers on RII11 (FLORDG032 and FLORDG033). This should be investigated.

    By Lori Garzio, on 3/21/19

    GI01SUMO-RII11-02-FLORDG031
    Deployment: 3

    It appears that the instrument failed shortly after deployment (values are negative). Data should be annotated with reason for failure (if known).

    By Lori Garzio, on 3/21/19

    CP02PMCO-WFP01-03-CTDPFK000
    Deployment: 1
    9/26/14, 12:00 AM 9/27/14, 12:00 AM

    suspect data with individual profiles

    By Leila Belabbassi, on 3/20/19

    CP02PMCO-WFP01-03-CTDPFK000
    Deployment: 6
    9/4/16, 12:00 AM 9/10/16, 12:00 AM

    suspect data with individual profiles

    By Leila Belabbassi, on 3/19/19

    CP02PMCO-WFP01-03-CTDPFK000
    Deployment: 8

    suspect data profiles

    By Leila Belabbassi, on 3/20/19