Single Point Velocity Meter

Reference Designator
CE02SHSM-RID26-04-VELPTA000
Review Status
Review Complete
Note
Depth
7m
Class
VELPT (Single Point Velocity Meter)
Make / Model
Nortek / Aquadopp 300m

Dataset Reviews Last processed: 8/7/19, 6:24 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_inst 174 174 0 0 0 0 16,559 900 7 / 8 Complete
2 recovered_inst 223 223 0 0 0 0 21,312 900 7 / 8 Complete
3 recovered_inst 135 135 0 0 0 0 12,842 900 7 / 8 Complete
4 recovered_inst 209 207 0 0 2 4 19,189 900 7 / 8 Complete
5 recovered_inst 179 179 0 0 0 0 17,095 900 7 / 7 Complete
6 recovered_inst 177 177 0 0 0 0 16,942 900 7 / 8 Complete
7 recovered_inst 176 176 0 0 0 0 16,863 900 7 / 8 Complete
Data Ranges Review Images

Test Notes

All good!

Data Coverage

Deployment: 1234567
100%100%100%99%100%100%100%

Lat/Lon Differences (km)

Deployment: 1234567
1 0.00
2 0.330.00
3 0.290.090.00
4 0.110.440.400.00
5 0.280.090.010.390.00
6 0.110.440.400.000.390.00
7 0.290.090.000.400.010.400.00

Example Composite Data Plot

System Annotations

Metadata Start Date End Date Comment
CE02SHSM-RID26-04-VELPTA000
4/2/15, 4:15 PM

According to the manufacturer, data are suspect when the instrument is tilted more than 20 degrees (measured by pitch and roll).

Id: 798 By: lgarzio

CE02SHSM-RID26
5/8/17, 8:00 PM 5/13/17, 8:00 PM

Communications with instruments on the NSIF for CE02SHSM are offline. Monitoring current draws to the NSIF shows a pattern of increased and expected current draws every 15 minutes, corresponding to the instrument sampling schedules. This suggests that the DCLs and attached instruments are still operating, but communications to the system are offline, reason unknown. Affected systems are: CPM2, DCL26: ADCPT, VELPT, PHSEN, uCSPP (acoustic modem communications), NUTNR, SPKIR, DCL27: OPTAA, FLORT, DOSTA, CTDBP. NSIF communicated starting again early on the 14th. Logs from NSIF systems indicate that the systems and instruments are still working, pointing to a communications issue between the NSIF and the surface buoy telemetry system.

Id: 865 By: michaesm
Flag: not_available Exclude: No

CE02SHSM-RID26
Method: telemetered
11/22/19, 3:45 PM 1/7/20, 1:30 PM

Due to reduced power on the mooring (limited recharging of the batteries with light winds and sea lions blocking the solar panels), we have had to reduce the operating schedule of the 7 meter platform. Normally we run the system continuously, but have reduced that schedule to hourly.

Id: 1835 By: wingardc

CE02SHSM-RID26
Method: telemetered
1/7/20, 7:00 PM 1/13/20, 1:45 PM

Due to an intermittent communications issue with DCL26, we are not receiving data from the instruments consistently. We expect the data to be available upon mooring recovery.

Id: 1890 By: wingardc
Flag: not_available Exclude: No

CE02SHSM-RID26
Method: telemetered
1/13/20, 1:45 PM

During a large storm event on 2020-01-11, power from the wind turbines was lost. This is the primary means of recharging the battery system on the mooring. With the loss of recharging power, we have had to implement a conservative power model. These systems and instruments have been disabled for the foreseeable future, and no further data will collected from them.

Id: 1902 By: wingardc
Flag: not_operational Exclude: No

Review Notes

Metadata Start Date End Date Comment
CE02SHSM-RID26-04-VELPTA000

15-29% of the data points collected during all deployments are when the instrument is tilted too far ( has pitch or roll >20 degrees). Example plot attached

By Lori Garzio, on 8/14/19

CE02SHSM-RID26-04-VELPTA000
Deployment: 6
10/10/17, 2:21 AM 10/20/17, 12:00 AM

Suspect data at the beginning of deployment 6.

By Lori Garzio, on 8/14/19

New Note