CTD

Reference Designator
CE04OSPS-SF01B-2A-CTDPFA107
Review Status
Review Complete
Note
Depth
20 to 200m
Class
CTDPF (CTD Profiler)
Make / Model
Sea-Bird / SBE 16plusV2

Dataset Reviews Last processed: 10/3/19, 4:31 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 streamed ctdpf_sbe43_sample 338 270 66 0 0 0 20,303 / 195 1 2 2 Complete
2 streamed ctdpf_sbe43_sample 351 344 0 1 2 6 116,227,103 / 220 1 2 2 Complete
3 streamed ctdpf_sbe43_sample 375 373 0 0 2 2 31,956,869 1 / 200 1 2 2 Complete
4 streamed ctdpf_sbe43_sample 354 46 3 304 1 1 3,811,656 1 / 291 1 2 2 Complete
Data Ranges Review Images

Test Notes

  1. unique and ascending not tested
  2. no other streams for comparison

Data Coverage

Deployment: 1234
ctdpf_sbe43_sample 80%98%99%13%

Lat/Lon Differences (km)

Deployment: 1234
1 0.00
2 0.080.00
3 0.080.000.00
4 0.080.000.000.00

System Annotations

Metadata Start Date End Date Comment
CE04OSPS
12/7/14, 2:45 PM 12/15/14, 7:00 PM

Note: PFE down. HVPS1 MOV explosion, 800A breaker tripped, investigation and restoration

Id: 1033 By: knuth

CE04OSPS
1/7/15, 2:32 AM 1/7/15, 3:06 AM

Note: PNWGP Portland <-> Seattle outage

Id: 1034 By: knuth

CE04OSPS
1/30/15, 7:00 PM 2/3/15, 7:00 PM

Note: Intermittent partial data loss due to storage drive problems at OTB

Id: 1035 By: knuth

CE04OSPS
3/21/15, 10:10 AM 3/22/15, 12:20 AM

Note: Intermittent data gaps. PNWGP outage due to City of Seattle fiber cable work

Id: 1037 By: knuth

CE04OSPS
6/12/15, 8:00 PM 6/15/15, 12:30 PM

Note: Intermittent data gaps. Network issues due to fire that damaged fibers between Portland and Seattle

Id: 1038 By: knuth

CE04OSPS-SF01B-2A-CTDPFA107
Method: streamed
Stream: ctdpf_sbe43_sample
8/2/15, 10:21 PM 7/17/16, 8:00 PM

Parameters: corrected_dissolved_oxygen, ext_volt0. Suspect values due to sensor malfunction.

Id: 1040 By: wruef
Flag: suspect Exclude: No

CE04OSPS
8/28/15, 8:00 PM 8/28/15, 8:30 PM

Note: Intermittent data gaps. Outage during major utility power failure in Seattle

Id: 1042 By: knuth

CE04OSPS
1/7/16, 1:10 AM 1/7/16, 1:52 AM

Note: Four 1-minute outages between Portland and Seattle due to maintenance

Id: 1043 By: knuth

CE04OSPS
3/10/16, 6:06 PM 3/11/16, 4:30 AM

Note: Fiber break between Portland and Seattle

Id: 1044 By: knuth

CE04OSPS-SF01B-2A-CTDPFA107
Method: streamed
Stream: ctdpf_sbe43_sample
4/14/16, 3:55 PM 4/19/16, 7:05 PM

Note: Data gap under investigation

Id: 1045 By: knuth
Flag: not_available Exclude: No

CE04OSPS-SF01B-2A-CTDPFA107
Method: streamed
Stream: ctdpf_sbe43_sample
4/23/16, 11:10 AM 4/24/16, 10:23 PM

Note: Data gap under investigation

Id: 1046 By: knuth
Flag: not_available Exclude: No

CE04OSPS
5/20/16, 12:33 PM 5/20/16, 2:04 PM

Note: Fiber break between Portland and Pacific City

Id: 1047 By: knuth

CE04OSPS
7/11/16, 10:53 PM 7/11/16, 11:51 PM

Note: Unexplained loss of power at Pittock Building in Portland

Id: 1048 By: knuth

CE04OSPS-SF01B-2A-CTDPFA107
7/12/16, 8:00 PM 7/19/16, 8:00 PM

Note: Intermittant data gaps as instrument comes back online.

Id: 1049 By: knuth

CE04OSPS-SF01B-2A-CTDPFA107
Method: streamed
Stream: ctdpf_sbe43_sample
7/18/16, 8:00 PM 7/27/17, 7:31 PM

Parameters: corrected_dissolved_oxygen, ext_volt0. Suspect values due to possible sensor malfunction.

Id: 1063 By: wruef
Flag: suspect Exclude: No

CE04OSPS-SF01B-2A-CTDPFA107
Method: streamed
Stream: ctdpf_sbe43_sample
7/19/16, 4:00 AM 7/27/17, 8:00 PM

Due to errors during vendor calibration, a calibration offset (CC_offset) of 9.6 is being applied during the on-demand creation of the L1 Seawater Pressure (dbar) data product from this instrument. For more information please contact help@oceanobservatories.org.

Id: 207 By: friedrich.knuth

CE04OSPS
12/17/16, 1:00 PM 12/17/16, 2:00 PM

Note: Corvalis data center lost power

Id: 1050 By: knuth

CE04OSPS
12/21/16, 8:50 PM 12/23/16, 7:44 AM

Note: Fiber break in Portland due to train crash

Id: 1051 By: knuth

CE04OSPS
1/8/17, 2:58 PM 1/8/17, 4:41 PM

Note: Network outage during major Seattle utility power failure

Id: 1052 By: knuth

CE04OSPS
1/9/17, 1:30 PM 1/9/17, 3:25 PM

Note: Lightning strike in Pacific City led to isolation of both cable lines from shore station equipment, when Power Feed Equipment High Voltage Opening Switches operated. All deployed infrastructure went offline.

Id: 1053 By: knuth

CE04OSPS-SF01B-2A-CTDPFA107
Method: streamed
Stream: ctdpf_sbe43_sample
1/9/17, 3:25 PM 1/10/17, 8:09 PM

CTD offline because of power interruption caused by lightning strike in Pacific City in close proximity to shore station.

Id: 1054 By: knuth
Flag: not_available Exclude: No

CE04OSPS-SF01B
1/30/17, 6:31 PM 7/27/17, 7:31 PM

Note: Due to the observed damage to the EM cable, the winched profiler has been returned to the dock and secured. Detailed inspection of engineering video of the Shallow Profiler-Endurance EM cable recorded earlier in the deployment indicates the anomalies most likely occurred in October 2016. Based on the video, it appears only the outer strength member is compromised. Due to the possibility the damage will continue to get worse and not wanting to risk the Science Pod (SciP) breaking free, profiling at this site has been restricted; however the SciP remains powered with its instruments active and collecting/transmitting data.

Id: 1056 By: knuth

CE04OSPS
2/7/17, 8:00 AM 2/7/17, 10:00 AM

Note: Intermittent data gaps. Outage during PNWGP 1-hour router-maintenance

Id: 1057 By: knuth

CE04OSPS
2/15/17, 9:43 AM 2/15/17, 12:16 PM

Note: Power Feed Equipment failure during major utility power issues related to lightning strike on Pacific City electrical sub-station. All deployed infrastructure went offline. Power was restored on February 16. No hardware issue subsea was identified.

Id: 1058 By: knuth

CE04OSPS-SF01B-2A-CTDPFA107
Method: streamed
Stream: ctdpf_sbe43_sample
2/15/17, 9:43 AM 2/15/17, 12:16 PM

Power Feed Equipment failure during major utility power issues related to lightning strike on Pacific City electrical sub-station. All deployed infrastructure went offline. Power was restored on February 16. No hardware issue subsea was identified.

Id: 1059 By: knuth
Flag: not_available Exclude: No

CE04OSPS-SF01B-2A-CTDPFA107
Method: streamed
Stream: ctdpf_sbe43_sample
2/15/17, 9:43 AM 2/16/17, 6:31 PM

CTD offline because of power interruption caused by lightning strike in Pacific City in close proximity to shore station.

Id: 1060 By: knuth
Flag: not_available Exclude: No

CE04OSPS-SF01B-2A-CTDPFA107
Method: streamed
Stream: ctdpf_sbe43_sample
8/1/17, 8:00 PM 9/16/17, 8:13 AM

Parameter: corrected_dissolved_oxygen. Suspect values due to possible issue with calibration data. Awaiting feedback from Marine Implementing Organization. Redmine 8668

Id: 1064 By: knuth
Flag: suspect Exclude: No

CE04OSPS
9/16/17, 7:30 AM 9/16/17, 8:13 AM

Data are suspect during this time range, due to the following event: On Saturday, September 16, 2017 at around 12:15 GMT, the CA Shallow Profiler at Endurance (CE04OSPS) went offline due to an overcurrent power event. Initial investigation indicated external aggression on one of the two anchor legs, pulling the platform that is normally fixed at 200m below the surface to 400m, and then rose to 100m before communications were lost. No data products are expected from CE04OSPS until the platform is re-deployed. For more information, please contact help@oceanobservatories.org

Id: 181 By: friedrich.knuth
Flag: suspect Exclude: No

CE04OSPS
Method: streamed
9/16/17, 8:13 AM 7/17/18, 11:00 AM

On Saturday, September 16, 2017 at around 12:15 GMT, the CA Shallow Profiler at Endurance (CE04OSPS) went offline due to an overcurrent power event. Initial investigation indicated external aggression on one of the two anchor legs, pulling the platform that is normally fixed at 200m below the surface to 400m, and then rose to 100m before communications were lost. No data products are expected from CE04OSPS until the platform is re-deployed. For more information, please contact help@oceanobservatories.org

Id: 221 By: friedrich.knuth
Flag: not_operational Exclude: No

CE04OSPS
8/22/18, 1:00 PM 8/24/18, 4:00 PM

On Wednesday Aug 22 2018, an automatic protective shutdown of primary power to the OOI Cabled Array occurred during installation of a new chiller system. Power was restored on Aug 24. Due to loss of power to the Cabled Array marine infrastructure, data were not collected by the instruments.

Id: 1485 By: lgarzio
Flag: not_operational Exclude: No

Review Notes

Metadata Start Date End Date Comment
CE04OSPS-SF01B

Annotation ID 181 is a duplicate of 221, with a much shorter time range.

By Lori Garzio, on 11/19/19

CE04OSPS-SF01B

There are no deployment depths in asset management for any of the instruments on the shallow profiler.

By Lori Garzio, on 11/19/19

CE04OSPS-SF01B-2A-CTDPFA107

Annotation IDs 1045 and 1046 from 2016 should be updated: "Note: Data gap under investigation"

By Lori Garzio, on 11/25/19

CE04OSPS-SF01B-2A-CTDPFA107

Annotation ID 1064 should be updated: "Parameter: corrected_dissolved_oxygen. Suspect values due to possible issue with calibration data. Awaiting feedback from Marine Implementing Organization. Redmine 8668"

By Lori Garzio, on 11/25/19

CE04OSPS-SF01B

There is a lot of missing data for deployment 1 for all instruments on the profiler. This should be annotated.

By Lori Garzio, on 11/25/19

CE04OSPS-SF01B-2A-CTDPFA107
Deployment: 2

Suspect data - particularly towards the end of the deployment (uniform density throughout the water column). This should be investigated and annotated.

By Lori Garzio, on 11/25/19

New Note