3-Wavelength Fluorometer

Reference Designator
CE05MOAS-GL327-02-FLORTM000
Review Status
Review Complete
Note
Depth
0 to 1,000m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBBCD-SLK

Dataset Reviews Last processed: 8/23/19, 7:18 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host flort_m_sample 80 79 0 0 0 0 2,773,700 1 200 / 196 Complete
2 recovered_host flort_m_sample 66 61 0 3 1 2 2,245,256 1 200 / 197 Complete
3 recovered_host flort_m_sample 33 32 0 0 0 0 1,044,298 1 200 / 198 Complete
4 recovered_host flort_m_sample 84 84 0 0 0 0 3,111,535 1 200 / 197 Complete
5 recovered_host flort_m_sample 85 85 0 0 0 0 2,730,609 1 1000 / 999 Complete
6 recovered_host flort_m_sample 63 61 0 1 1 1 2,510,682 1 200 / 203 Complete
7 recovered_host flort_m_sample 86 86 0 0 0 0 3,530,193 1 1000 / 998 Complete
Data Ranges Review Images

Test Notes

All good!

Data Coverage

Deployment: 1234567
flort_m_sample 99%92%97%100%100%97%100%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
CE05MOAS-GL327
12/11/15, 7:00 PM 12/19/15, 5:00 AM

Glider leaked on 2015-12-12 and was set to drift until recovery on 2015-12-19. On 2015-12-15, a test dive was performed that resulted in another abort due to a leak, but provided a small addition of data in the recovered set.

Id: 1469 By: michaesm

CE05MOAS-GL327-02-FLORTM000
3/1/17, 7:00 PM 3/7/17, 7:00 PM

Instrument was turned off on 2017-03-02 to prolong battery life to recovery date.

Id: 892 By: michaesm
Flag: not_operational Exclude: No

Review Notes

Metadata Start Date End Date Comment
CE05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CE05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of 3 consecutive down and up casts for CE05MOAS-GL312-04-DOSTAM000 on 7/8/2014 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/14/19

CE05MOAS-GL327-02-FLORTM000
Deployment: 1

Suspect profiles starting halfway through the deployment.

By Lori Garzio, on 8/28/19

CE05MOAS-GL327-02-FLORTM000
Deployment: 2

Suspect data for the 2nd half of the deployment

By Lori Garzio, on 8/28/19

CE05MOAS-GL327-02-FLORTM000
Deployment: 4
Status: Unavailable
3/2/17, 12:00 AM 3/8/17, 12:00 AM

The PAR (PARAD) and ECO triplet (FLORT) was later turned off on 2017-03-02 to further prolong battery life to recovery date.

By Mike Smith, on 3/15/17

New Note