3-Wavelength Fluorometer

Reference Designator
CE05MOAS-GL383-02-FLORTM000
Review Status
Review Complete
Note
Depth
0 to 1,000m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBBCD-SLK

Dataset Reviews Last processed: 4/10/19, 3:42 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host flort_m_sample 52 50 0 1 0 0 1,877,570 1 1000 / 996 1 Complete
2 recovered_host flort_m_sample 105 104 0 0 0 0 2,984,339 1 200 / 196 Complete
3 recovered_host flort_m_sample 7 5 0 1 0 0 126,072 1 200 / 150 Complete
4 recovered_host flort_m_sample 100 100 0 0 0 0 2,569,210 1 200 / 195 Complete
5 recovered_host flort_m_sample 83 83 0 0 0 0 3,361,242 1 200 / 201 Complete
Data Ranges Review Images

Test Notes

  1. fail: data found in another stream (gaps: [1] days: [2])

Data Coverage

Deployment: 12345
flort_m_sample 96%99%71%100%100%

Deployment Map

System Annotations

No annotations yet.

Review Notes

Metadata Start Date End Date Comment
CE05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CE05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of 3 consecutive down and up casts for CE05MOAS-GL312-04-DOSTAM000 on 7/8/2014 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/14/19

CE05MOAS-GL383-02-FLORTM000
Deployment: 1

There are telemetered data available for 2015-02-27T19:29:50 to 2015-02-28T03:03:13, but no recovered_host data are available. These data should be available in the recovered_host data stream.

By Lori Garzio, on 5/3/19

CE05MOAS-GL383-02-FLORTM000
Deployment: 1

Suspect profile on/around 3/1/2015.

By Lori Garzio, on 5/7/19

CE05MOAS-GL383-02-FLORTM000
Deployment: 2

Suspect data on/around 9/15/2015

By Lori Garzio, on 5/7/19

CE05MOAS-GL383
Deployment: 3

Data should be annotated to explain why the deployment was so short.

By Lori Garzio, on 5/1/19

CE05MOAS-GL383-02-FLORTM000
Deployment: 4

Suspect backscatter data at the end of the deployment.

By Lori Garzio, on 5/7/19

CE05MOAS-GL383
Deployment: 4
3/1/17, 12:00 AM 4/10/17, 12:00 AM

This glider looks like it had a sampling change (similar to glider 319, see annotation ID 1371). This dataset should be annotated.

By Lori Garzio, on 5/1/19

CE05MOAS-GL383-02-FLORTM000
Deployment: 5

Suspect backscatter data near the end of the deployment

By Lori Garzio, on 5/7/19

New Note