Bio-acoustic Sonar (Coastal)

Reference Designator
CP01CNSM-MFD37-07-ZPLSCC000
Review Status
Will Not Review
Note
Not Delivered Via uFrame
Depth
133m
Class
ZPLSC (Bio-acoustic Sonar (Coastal))
Make / Model
ASL Environmental Sciences / AZFP

Dataset Reviews Last processed:

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
3 telemetered zplsc_c_instrument / Todo
5 telemetered zplsc_c_instrument / Todo
6 telemetered zplsc_c_instrument / Todo
7 telemetered zplsc_c_instrument / Todo
8 telemetered zplsc_c_instrument / Todo
Data Ranges Review Images

Test Notes

All good!

System Annotations

Metadata Start Date End Date Comment
CP01CNSM
2/16/14, 7:00 PM 4/17/14, 8:00 PM

Surface buoy exploded due to hydrogen build-up in well.

Id: 103 By: swhite
Flag: not_operational Exclude: No

CP01CNSM
12/13/14, 1:47 PM 12/15/14, 3:58 PM

Deployment 2: Short deployment due to the fuel cell not functioning correctly. No data expected.

Id: 709 By: cdobson

CP01CNSM-MFD37-07-ZPLSCC000
5/7/15, 1:34 PM 10/23/15, 3:40 PM

Recovered data will be made available when the raw data parser is ready.

Id: 541 By: leila

CP01CNSM-MFD37-07-ZPLSCC000
Stream: zplsc_c_instrument
5/7/15, 1:34 PM

ZPLSC data may contain acoustic interference from the ADCP instrument subject to real-time clock drift of either instrument.

Id: 2005 By: swhite

CP01CNSM-MFD37-07-ZPLSCC000
5/13/16, 9:50 AM 10/13/16, 3:34 PM

Recovered data will be made available when the raw data parser is ready.

Id: 545 By: leila

CP01CNSM-MFD37
5/21/16, 8:00 PM 5/31/16, 8:02 AM

Issue observed with the Seafloor Multi-Function Node may have caused loss of data.

Id: 573 By: leila

CP01CNSM-MFD37-07-ZPLSCC000
10/13/16, 2:36 PM 6/9/17, 12:05 PM

Recovered data will be made available when the raw data parser is ready.

Id: 549 By: leila

CP01CNSM-MFD37-07-ZPLSCC000
10/13/16, 2:36 PM 6/9/17, 12:05 PM

Deployment 6: 200 kHz transducer failure; Post-recovery evaluation of the instrument revealed failures or defects affecting the 200 kHz transducer. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.

Id: 2023 By: swhite
Flag: suspect Exclude: No

CP01CNSM-MFD37
1/21/17, 7:00 PM 6/9/17, 11:00 AM

Multi-function Node was shut down due to low power.

Id: 105 By: lgarzio
Flag: not_operational Exclude: No

CP01CNSM-MFD37-07-ZPLSCC000
6/9/17, 10:24 AM 11/1/17, 4:33 PM

Recovered data will be made available when the raw data parser is ready.

Id: 550 By: leila

CP01CNSM-MFD37
12/20/17, 3:00 AM 3/24/18, 6:00 PM

Multi-Function Node was disabled due to shorting in the power.

Id: 337 By: leila
Flag: not_operational Exclude: No

CP01CNSM-MFD37-07-ZPLSCC000
3/24/18, 5:32 PM 10/29/18, 8:31 AM

Deployment 9: Multiple transducer failures; Post-recovery evaluation of the instrument revealed failures or defects affecting the 38 kHz and 125 kHz transducers. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.

Id: 2024 By: swhite
Flag: suspect Exclude: No

CP01CNSM-MFD37
Method: telemetered
3/27/18, 11:03 PM 10/29/18, 9:48 PM

Deployment 9: An unexpected ground fault had occurred. Instruments with batteries will continue to collect data.

Id: 1364 By: leila
Flag: not_operational Exclude: No

CP01CNSM-MFD37
Method: telemetered
10/27/18, 9:00 AM 4/7/19, 2:08 PM

Deployment 10: Power to MFN was lost immediately after deployment. There will be no telemetered data for any MFN instruments for this deployment. Most impacted instruments will continue to collect data on battery power and be available as recovered data.

Id: 1558 By: cdobson
Flag: not_operational Exclude: No

CP01CNSM-MFD37-07-ZPLSCC000
10/29/18, 9:48 PM 4/7/19, 2:08 PM

Deployment 10: 200 kHz transducer failure; Post-recovery evaluation of the instrument revealed failures or defects affecting the 200 kHz transducer. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.

Id: 2027 By: swhite
Flag: suspect Exclude: No

CP01CNSM-MFD37-07-ZPLSCC000
4/6/19, 10:35 AM 9/26/19, 1:15 PM

Deployment 11: Multiple transducer failures; Post-recovery evaluation of the instrument revealed failures or defects affecting the 38 kHz and 125 kHz transducers. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.

Id: 2029 By: swhite
Flag: suspect Exclude: No

CP01CNSM-MFD37
9/27/19, 2:00 PM

Deployment 12: Power issue hinders telemtered data transfer from all MFN instruments. Data from all internally logging instruments will be available on recovery.

Id: 1742 By: cdobson

Review Notes

Metadata Start Date End Date Comment
CP01CNSM-MFD37

Annotation IDs 1564 and 1565 are duplicates of annotation ID 1558.

By Lori Garzio, on 3/25/19

CP01CNSM
Deployment: 2

Annotation ID 709 "Short deployment due to the fuel cell not functioning correctly" should be updated to indicate that this refers to deployment 2, and data aren't expected from most instruments for this time period.

By Lori Garzio, on 3/22/19

New Note