Method | Data Stream | Content | Type | |
---|---|---|---|---|
recovered_host | zplsc_c_instrument | Data Products | Report M2M Stats | Science |
recovered_inst | zplsc_echogram_data | Data Products | Report M2M Stats | Science |
telemetered | zplsc_c_instrument | Data Products | Report M2M Stats | Science |
Deployment | Cruise | Start Date | Stop Date | Mooring Asset | Node Asset | Sensor Asset | Latitude | Longitude | Deployment Depth | Water Depth |
---|---|---|---|---|---|---|---|---|---|---|
3 Review | AT-27 | 05/07/2015 | 10/23/2015 | CGMCP-01CNSM-00003 | CGINS-ZPLSCC-55080 | 40.1402 | -70.7713 | 134 | 135 | |
5 Review | AR-04 | 05/13/2016 | 10/13/2016 | CGMCP-01CNSM-00005 | CGINS-ZPLSCC-55074 | 40.1404 | -70.7713 | 133 | 134 | |
6 Review | AR-08 | 10/13/2016 | 06/09/2017 | CGMCP-01CNSM-00006 | CGINS-ZPLSCC-55102 | 40.1334 | -70.7785 | 133 | 134 | |
7 Review | AR-18 | 06/09/2017 | 11/01/2017 | CGMCP-01CNSM-00007 | CGINS-ZPLSCC-55077 | 40.1398 | -70.7711 | 133 | 134 | |
8 Review | AR-24 | 10/29/2017 | 03/29/2018 | CGMCP-01CNSM-00008 | CGINS-ZPLSCC-55102 | 40.1334 | -70.7783 | 133 | 134 | |
9 Review | AR-28 | 03/24/2018 | 10/29/2018 | CGMCP-01CNSM-00009 | CGINS-ZPLSCC-55100 | 40.1398 | -70.7713 | 133 | 134 | |
10 Review | AR-31 | 10/30/2018 | 04/07/2019 | CGMCP-01CNSM-00010 | CGINS-ZPLSCC-55102 | 40.1334 | -70.7777 | 133 | 134 | |
11 Review | AR-34 | 04/06/2019 | 09/26/2019 | CGMCP-01CNSM-00011 | CGINS-ZPLSCC-55101 | 40.1401 | -70.7714 | 132 | 133 | |
12 Review | AR-39 | 09/27/2019 | CGMCP-01CNSM-00012 | CGINS-ZPLSCC-55080 | 40.1332 | -70.7783 | 132 | 133 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP01CNSM-MFD37 |
Annotation IDs 1564 and 1565 are duplicates of annotation ID 1558. By Lori Garzio, on 3/25/19 |
||
CP01CNSM Deployment: 2 |
Annotation ID 709 "Short deployment due to the fuel cell not functioning correctly" should be updated to indicate that this refers to deployment 2, and data aren't expected from most instruments for this time period. By Lori Garzio, on 3/22/19 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP01CNSM |
2/16/14, 7:00 PM | 4/17/14, 8:00 PM |
Surface buoy exploded due to hydrogen build-up in well.
Id: 103
By: swhite
|
CP01CNSM |
12/13/14, 1:47 PM | 12/15/14, 3:58 PM |
Deployment 2: Short deployment due to the fuel cell not functioning correctly. No data expected. Id: 709 By: cdobson |
CP01CNSM-MFD37-07-ZPLSCC000 |
5/7/15, 1:34 PM | 10/23/15, 3:40 PM |
Recovered data will be made available when the raw data parser is ready. Id: 541 By: leila |
CP01CNSM-MFD37-07-ZPLSCC000 Stream: zplsc_c_instrument |
5/7/15, 1:34 PM |
ZPLSC data may contain acoustic interference from the ADCP instrument subject to real-time clock drift of either instrument. Id: 2005 By: swhite |
|
CP01CNSM-MFD37-07-ZPLSCC000 |
5/13/16, 9:50 AM | 10/13/16, 3:34 PM |
Recovered data will be made available when the raw data parser is ready. Id: 545 By: leila |
CP01CNSM-MFD37 |
5/21/16, 8:00 PM | 5/31/16, 8:02 AM |
Issue observed with the Seafloor Multi-Function Node may have caused loss of data. Id: 573 By: leila |
CP01CNSM-MFD37-07-ZPLSCC000 |
10/13/16, 2:36 PM | 6/9/17, 12:05 PM |
Recovered data will be made available when the raw data parser is ready. Id: 549 By: leila |
CP01CNSM-MFD37-07-ZPLSCC000 |
10/13/16, 2:36 PM | 6/9/17, 12:05 PM |
Deployment 6: 200 kHz transducer failure; Post-recovery evaluation of the instrument revealed failures or defects affecting the 200 kHz transducer. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.
Id: 2023
By: swhite
|
CP01CNSM-MFD37 |
1/21/17, 7:00 PM | 6/9/17, 11:00 AM |
Multi-function Node was shut down due to low power.
Id: 105
By: lgarzio
|
CP01CNSM-MFD37-07-ZPLSCC000 |
6/9/17, 10:24 AM | 11/1/17, 4:33 PM |
Recovered data will be made available when the raw data parser is ready. Id: 550 By: leila |
CP01CNSM-MFD37 |
12/20/17, 3:00 AM | 3/24/18, 6:00 PM |
Multi-Function Node was disabled due to shorting in the power.
Id: 337
By: leila
|
CP01CNSM-MFD37-07-ZPLSCC000 |
3/24/18, 5:32 PM | 10/29/18, 8:31 AM |
Deployment 9: Multiple transducer failures; Post-recovery evaluation of the instrument revealed failures or defects affecting the 38 kHz and 125 kHz transducers. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.
Id: 2024
By: swhite
|
CP01CNSM-MFD37 Method: telemetered |
3/27/18, 11:03 PM | 10/29/18, 9:48 PM |
Deployment 9: An unexpected ground fault had occurred. Instruments with batteries will continue to collect data.
Id: 1364
By: leila
|
CP01CNSM-MFD37 Method: telemetered |
10/27/18, 9:00 AM | 4/7/19, 2:08 PM |
Deployment 10: Power to MFN was lost immediately after deployment. There will be no telemetered data for any MFN instruments for this deployment. Most impacted instruments will continue to collect data on battery power and be available as recovered data.
Id: 1558
By: cdobson
|
CP01CNSM-MFD37-07-ZPLSCC000 |
10/29/18, 9:48 PM | 4/7/19, 2:08 PM |
Deployment 10: 200 kHz transducer failure; Post-recovery evaluation of the instrument revealed failures or defects affecting the 200 kHz transducer. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.
Id: 2027
By: swhite
|
CP01CNSM-MFD37-07-ZPLSCC000 |
4/6/19, 10:35 AM | 9/26/19, 1:15 PM |
Deployment 11: Multiple transducer failures; Post-recovery evaluation of the instrument revealed failures or defects affecting the 38 kHz and 125 kHz transducers. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.
Id: 2029
By: swhite
|
CP01CNSM-MFD37 |
9/27/19, 2:00 PM |
Deployment 12: Power issue hinders telemtered data transfer from all MFN instruments. Data from all internally logging instruments will be available on recovery. Id: 1742 By: cdobson |