Spectral Irradiance

Reference Designator
CP01CNSM-RID26-08-SPKIRB000
Review Status
Review Complete
Note
Depth
7m
Class
SPKIR (Spectral Irradiance)
Make / Model
Satlantic / OCR507 ICSW

Dataset Reviews Last processed: 7/9/19, 8:09 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 telemetered 149 34 0 114 0 0 1,008,823 1 7 / 1 1 2 Complete
2 telemetered 3 3 0 0 0 0 22,925 1 7 / 1 1 2 Complete
3 recovered_host 170 170 0 0 0 0 2,393,786 1 7 / 3 2 Complete
4 recovered_host 165 163 0 2 0 0 2,476,285 1 7 / 1 1 2 Complete
5 recovered_host 154 139 0 0 5 19 2,190,042 1 7 / 3 2 Complete
6 recovered_host 240 240 0 0 0 0 3,381,870 1 7 / 3 2 Complete
7 recovered_host 146 131 0 1 1 14 2,087,107 1 7 / 3 2 Complete
8 recovered_host 152 152 0 0 0 0 2,487,750 1 7 / 3 2 Complete
Data Ranges Review Images

Test Notes

  1. no other streams for comparison
  2. missing: ['pressure']
  3. no comparison: timestamps do not match

Data Coverage

Deployment: 12345678
23%100%
100%99%90%100%90%100%

Lat/Lon Differences (km)

Deployment: 12345678
1 0.00
2 0.840.00
3 0.400.950.00
4 0.830.090.980.00
5 0.420.970.021.000.00
6 0.830.070.970.020.990.00
7 0.350.930.050.960.070.950.00
8 0.820.080.960.020.980.020.940.00

Example Composite Data Plot

System Annotations

Metadata Start Date End Date Comment
CP01CNSM-RID26-08-SPKIRB000
11/21/13, 1:15 PM 4/17/14, 8:00 PM

Data not recoverable.

Id: 504 By: leila

CP01CNSM
2/16/14, 7:00 PM 4/17/14, 8:00 PM

Surface buoy exploded due to hydrogen build-up in well.

Id: 103 By: swhite
Flag: not_operational Exclude: No

CP01CNSM-RID26-08-SPKIRB000
12/13/14, 1:47 PM 12/15/14, 3:58 PM

Data not recoverable.

Id: 519 By: leila

CP01CNSM
12/13/14, 1:47 PM 12/15/14, 3:58 PM

Deployment 2: Short deployment due to the fuel cell not functioning correctly. No data expected.

Id: 709 By: cdobson

CP01CNSM-RID26-08-SPKIRB000
Method: telemetered
10/23/15, 2:49 PM 4/4/16, 8:03 AM

Data not telemetered.

Id: 363 By: leila
Flag: not_available Exclude: No

CP01CNSM-RID26
Method: telemetered
4/23/18, 1:22 PM 10/29/18, 9:48 PM

Power down - Instrument Disabled.

Id: 1379 By: leila
Flag: not_operational Exclude: No

CP01CNSM-RID26-08-SPKIRB000
9/27/19, 2:30 PM

The use of UV lamps has been implemented for bio-fouling mitigation on this instrument as of the start date/time of this annotation.

Id: 1773 By: swhite

Review Notes

Metadata Start Date End Date Comment
CP01CNSM-RID26-08-SPKIRB000

There is no pressure variable in the .nc files. This should be a coordinate.

By Lori Garzio, on 7/12/19

CP01CNSM-RID26-08-SPKIRB000

A substantial percentage (between 9-32%) of data are outside of global ranges for every deployment. Zooming in on the x-axis in the plots shows that all of the values that are outside of global ranges are negative.

By Lori Garzio, on 7/12/19

CP01CNSM-RID26-08-SPKIRB000

missing data for deployments 5 and 7 should be annotated.

By Lori Garzio, on 7/12/19

CP01CNSM
Deployment: 2

Annotation ID 709 "Short deployment due to the fuel cell not functioning correctly" should be updated to indicate that this refers to deployment 2, and data aren't expected from most instruments for this time period.

By Lori Garzio, on 3/22/19

New Note