Photosynthetically Available Radiation

Reference Designator
CP04OSPM-WFP01-05-PARADK000
Review Status
Review Complete
Note
Depth
15 to 427m
Class
PARAD (Photosynthetically Available Radiation)
Make / Model
Biospherical Instruments / QSP-2200

Dataset Reviews Last processed: 8/22/19, 2:49 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_wfp 173 173 0 0 0 0 300,938 5 427 / 429 Complete
2 recovered_wfp 208 207 0 0 0 0 139,575 5 427 / 373 Complete
3 recovered_wfp 167 167 0 0 0 0 153,782 6 427 / 417 Complete
4 recovered_wfp 221 221 0 0 0 0 385,825 5 427 / 411 Complete
5 recovered_wfp 133 132 0 0 0 0 76,522 5 427 / 338 Complete
6 recovered_wfp 244 13 0 231 0 0 19,457 5 427 / 410 Complete
7 recovered_wfp 157 157 0 0 0 0 260,396 5 427 / 419 Complete
8 recovered_wfp 152 55 0 96 0 0 78,855 6 427 / 414 Complete
Data Ranges Review Images

Test Notes

All good!

Data Coverage

Deployment: 12345678
100%100%100%100%99%5%100%36%

Lat/Lon Differences (km)

Deployment: 12345678
1 0.00
2 0.830.00
3 0.580.620.00
4 0.640.670.940.00
5 0.590.730.111.020.00
6 0.590.580.850.100.930.00
7 0.580.730.111.010.010.920.00
8 0.670.771.020.101.100.201.090.00

System Annotations

Metadata Start Date End Date Comment
CP04OSPM-WFP01
11/2/14, 10:12 AM 4/28/15, 11:13 AM

Wire Following Profiler is stuck around 55 dbar.

Id: 1393 By: leila

CP04OSPM-WFP01
8/5/15, 8:14 AM 10/12/15, 5:05 AM

Wire Following Profiler is stuck at the bottom of its range.

Id: 1394 By: leila

CP04OSPM-WFP01
6/2/16, 11:00 AM 9/28/16, 9:25 AM

Wire Following Profiler is stuck at the top of its range.

Id: 613 By: leila

CP04OSPM
10/12/16, 2:00 AM 6/1/17, 7:22 PM

Failed

Id: 354 By: leila
Flag: not_operational Exclude: No

CP04OSPM-WFP01
1/2/18, 7:00 PM 4/9/18, 12:25 PM

Deployment 8: Wire-Following Profiler stopped transmitting data. Issue identified as a software problem.

Id: 227 By: leila
Flag: not_available Exclude: No

CP04OSPM-WFP01
12/31/19, 7:00 PM 6/8/20, 12:19 PM

Deployment 12: Due to a clock rollover issue, all WFPs have ceased sampling on 1/1/2020. No WFP data are being collected after this date. A resolution is being researched by the vendor.

Id: 1885 By: cdobson

Review Notes

Metadata Start Date End Date Comment
CP04OSPM-WFP01-05-PARADK000

51% - 98% of parad_k_par values for all deployments are outside of global ranges. Zooming in shows that the majority of the values outside of global ranges are negative. This should be investigated.

By Lori Garzio, on 8/26/19

CP04OSPM-WFP01
9/26/16, 12:00 AM 9/28/16, 1:25 PM

Deployment 5: suspect data at 0 dbar at the very end of the deployment

By Lori Garzio, on 8/26/19

New Note