Method | Data Stream | Content | Type | |
---|---|---|---|---|
recovered_wfp | parad_k__stc_imodem_instrument_recovered | Data Products | Report M2M Stats | Science |
telemetered | parad_k__stc_imodem_instrument | Data Products | Report M2M Stats | Science |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP04OSPM-WFP01-05-PARADK000 |
51% - 98% of parad_k_par values for all deployments are outside of global ranges. Zooming in shows that the majority of the values outside of global ranges are negative. This should be investigated. By Lori Garzio, on 8/26/19 |
||
CP04OSPM-WFP01 |
9/26/16, 12:00 AM | 9/28/16, 1:25 PM |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP04OSPM-WFP01 |
11/2/14, 10:12 AM | 4/28/15, 11:13 AM |
Wire Following Profiler is stuck around 55 dbar. Id: 1393 By: leila |
CP04OSPM-WFP01 |
8/5/15, 8:14 AM | 10/12/15, 5:05 AM |
Wire Following Profiler is stuck at the bottom of its range. Id: 1394 By: leila |
CP04OSPM-WFP01 |
6/2/16, 11:00 AM | 9/28/16, 9:25 AM |
Wire Following Profiler is stuck at the top of its range. Id: 613 By: leila |
CP04OSPM |
10/12/16, 2:00 AM | 6/1/17, 7:22 PM |
Failed
Id: 354
By: leila
|
CP04OSPM-WFP01 |
1/2/18, 7:00 PM | 4/9/18, 12:25 PM |
Deployment 8: Wire-Following Profiler stopped transmitting data. Issue identified as a software problem.
Id: 227
By: leila
|
CP04OSPM-WFP01 |
12/31/19, 7:00 PM | 6/8/20, 12:19 PM |
Deployment 12: Due to a clock rollover issue, all WFPs have ceased sampling on 1/1/2020. No WFP data are being collected after this date. A resolution is being researched by the vendor. Id: 1885 By: cdobson |