3-Wavelength Fluorometer

Reference Designator
CP05MOAS-GL374-02-FLORTM000
Review Status
Review Complete
Note
Depth
0 to 1,000m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBBCD-SLK

Dataset Reviews Last processed: 8/29/19, 3:26 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host flort_m_sample 87 86 0 0 0 0 1,906,836 1 200 / 158 1 Complete
2 recovered_host flort_m_sample 76 75 0 0 0 0 4,887,884 1 200 / 200 2 Complete
3 recovered_host flort_m_sample 73 72 0 0 0 0 4,935,809 1 200 / 196 3 Complete
4 recovered_host flort_m_sample 69 67 0 1 1 1 4,363,521 1 200 / 195 3 Complete
Data Ranges Review Images

Test Notes

  1. fail: data found in another stream (gaps: [10] days: [10])
  2. fail: data found in another stream (gaps: [11] days: [11])
  3. fail: data found in another stream (gaps: [5] days: [5])

Data Coverage

Deployment: 1234
flort_m_sample 99%99%99%97%

Deployment Map

System Annotations

No annotations yet.

Review Notes

Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-GL374-02-FLORTM000

Comparisons with the shipboard CTD casts were done for D0001, D0002, D0003 and D0004. Chl-a recorded by the glider was higher compared to the CTD cast (particularly at shallower depths). Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/13/19

CP05MOAS-GL374-02-FLORTM000
Deployment: 1

Suspect data towards the end of the deployment.

By Lori Garzio, on 9/3/19

CP05MOAS-GL374-02-FLORTM000
Deployment: 2

Suspect data particularly in the second half of the deployment, mostly >50 dbar

By Lori Garzio, on 9/3/19

CP05MOAS-GL374
Deployment: 4

Glider looks like it drifted at the end of the deployment - this should be annotated

By Lori Garzio, on 9/3/19