3-Wavelength Fluorometer

Reference Designator
CP05MOAS-GL374-02-FLORTM000
Start Depth
0
End Depth
1,000
Location
Current Status
Review Complete
Class
FLORT (3-Wavelength Fluorometer)
Series
FLORT-M
Science Discipline
Biological
Make
WET Labs
Model
ECO Puck FLBBCD-SLK
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CP05MOAS/GL374/02-FLORTM000/metadata
Method Data Stream Content Type
recovered_host flort_m_sample Data Products Report M2M Stats Science
telemetered flort_m_sample Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review KN-217 04/16/2014 07/11/2014 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00374 CGINS-FLORTM-02819 40.1115 -70.5555 200 200
2 Review AT-27 05/09/2015 07/24/2015 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00374 CGINS-FLORTM-02819 39.9418 -70.8266 200 200
3 Review AR-04 05/29/2016 08/09/2016 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00374 CGINS-FLORTM-04195 40.1195 -70.4483 200 200
4 Review AR-08 10/11/2016 12/20/2016 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00374 CGINS-FLORTM-04468 40.0873 -70.3782 200 200
5 Review SI20170418 04/20/2017 04/20/2017 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00374 CGINS-FLORTM-04468 40.0288 -70.4857 200 200
Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-GL374-02-FLORTM000

Comparisons with the shipboard CTD casts were done for D0001, D0002, D0003 and D0004. Chl-a recorded by the glider was higher compared to the CTD cast (particularly at shallower depths). Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/13/19

CP05MOAS-GL374-02-FLORTM000
Deployment: 1

Suspect data towards the end of the deployment.

By Lori Garzio, on 9/3/19

CP05MOAS-GL374-02-FLORTM000
Deployment: 2

Suspect data particularly in the second half of the deployment, mostly >50 dbar

By Lori Garzio, on 9/3/19

CP05MOAS-GL374
Deployment: 4

Glider looks like it drifted at the end of the deployment - this should be annotated

By Lori Garzio, on 9/3/19

New Note

No annotations yet.