Direct Covariance Flux

Reference Designator
CP01CNSM-SBD12-08-FDCHPA000
Start Depth
-3
End Depth
-3
Location
Buoy Tower
Current Status
Review Complete
Class
FDCHP (Direct Covariance Flux)
Series
FDCHP-A
Science Discipline
Physical
Make
WHOI
Model
DCFS
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CP01CNSM/SBD12/08-FDCHPA000/metadata
Method Data Stream Content Type
recovered_host fdchp_a_dcl_instrument_recovered Data Products Report M2M Stats Science
recovered_inst fdchp_a_instrument_recovered Data Products Report M2M Stats Science
telemetered fdchp_a_dcl_instrument Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
2 Review KN-224 12/13/2014 12/15/2014 CGMCP-01CNSM-00002 CGINS-FDCHPA-41005 40.1339 -70.7789 -5 135
3 Review AT-27 05/07/2015 10/23/2015 CGMCP-01CNSM-00003 CGINS-FDCHPA-41005 40.1402 -70.7713 -5 135
4 Review AT-31 10/23/2015 04/04/2016 CGMCP-01CNSM-00004 CGINS-FDCHPA-44904 40.1332 -70.7784 -5 136
5 Review AR-04 05/13/2016 10/13/2016 CGMCP-01CNSM-00005 CGINS-FDCHPA-44727 40.1404 -70.7713 -5 134
6 Review AR-08 10/13/2016 06/09/2017 CGMCP-01CNSM-00006 CGINS-FDCHPA-41005 40.1334 -70.7785 -5 134
7 Review AR-18 06/09/2017 11/01/2017 CGMCP-01CNSM-00007 CGINS-FDCHPA-44903 40.1398 -70.7711 -5 134
8 Review AR-24 10/29/2017 03/29/2018 CGMCP-01CNSM-00008 CGINS-FDCHPA-44727 40.1334 -70.7783 -5 134
9 Review AR-28 03/24/2018 10/29/2018 CGMCP-01CNSM-00009 CGINS-FDCHPA-44904 40.1398 -70.7713 -5 134
10 Review AR-31 10/30/2018 04/07/2019 CGMCP-01CNSM-00010 CGINS-FDCHPA-44727 40.1334 -70.7777 -5 134
11 Review AR-34 04/06/2019 09/26/2019 CGMCP-01CNSM-00011 CGINS-FDCHPA-64401 40.1401 -70.7714 -5 133
12 Review AR-39 09/27/2019 CGMCP-01CNSM-00012 CGINS-FDCHPA-44902 40.1332 -70.7783 -5 133
Metadata Start Date End Date Comment
CP01CNSM-SBD12-08-FDCHPA000

Re annotation ID 1374 for this instrument - the equivalent annotation on CE02SHSM-SBD12-08-FDCHPA000 (ID 1373) was updated by wingardc: "UPDATED, 2019-02-05: Per communication with the instrument subject matter expert, the onboard processing error applies only to the motion-corrected wind speed and flux calculations. Those parameters are marked here with a QC Flag = "fail". All other measurements are considered to be good. Users should only use the motion-corrected wind speed and flux data from the recovered instrument data stream." This information presumably also applies to this instrument and should be communicated to users.

By Lori Garzio, on 6/7/19

CP01CNSM-SBD12-08-FDCHPA000

There are no global ranges in the system for this instrument for fdchp_a_fluxhot, fdchp_a_fluxmom_alongwind, and fdchp_a_fluxmom_crosswind

By Lori Garzio, on 6/7/19

CP01CNSM-SBD12-08-FDCHPA000

Data arrays are all fill values for every deployment for fdchp_a_tmpatur, fdchp_a_windtur_north, fdchp_a_windtur_up, fdchp_a_windtur_west, fdchp_a_fluxhot, fdchp_a_fluxmom_alongwind, fdchp_a_fluxmom_crosswind (Redmine 13402). There are several errors listed in the computed_provenance section of the provenance.json files (one example file attached). When this issue is resolved annotation ID 1397 should be updated.

By Lori Garzio, on 6/7/19

CP01CNSM
Deployment: 2

Annotation ID 709 "Short deployment due to the fuel cell not functioning correctly" should be updated to indicate that this refers to deployment 2, and data aren't expected from most instruments for this time period.

By Lori Garzio, on 3/22/19

CP01CNSM-SBD12-08-FDCHPA000
Deployment: 4

Annotation ID 544: recovered data are missing. These data should be available for download or the annotation should be updated to explain why the data aren't available.

By Lori Garzio, on 6/7/19

New Note

Metadata Start Date End Date Comment
CP01CNSM
2/16/14, 7:00 PM 4/17/14, 8:00 PM

Surface buoy exploded due to hydrogen build-up in well.

Id: 103 By: swhite
Flag: not_operational Exclude: No

CP01CNSM-SBD12-08-FDCHPA000
12/13/14, 1:47 PM 10/12/16, 8:00 PM

Sonic temperature bugs: general slow response problem leading to errors of +/- 6 degrees from ambient, and an additional nonlinear error in the slow response when the ambient temperature is between -5 and 5 degrees. The fast response (> 1 Hz) of these sensors is still accurate. Therefore, flux measurements using the detrended temperature data should still be valid. w bug: positive vertical velocities w output by the anemometer need to be multiplied by 1.166; negative vertical velocities need to be multiplied by 1.289 (Redmine 11645)

Id: 80 By: lgarzio
Flag: suspect Exclude: No

CP01CNSM-SBD12-08-FDCHPA000
Method: telemetered
Stream: fdchp_a_dcl_instrument
12/13/14, 1:47 PM 9/26/19, 1:15 PM

UPDATED, 2019-11-07: Per communication with the instrument subject matter expert, the onboard processing error applies only to the motion-corrected wind speed and flux calculations. Those parameters are marked here with a QC Flag = "fail". All other measurements are considered to be good. Users should only use the motion-corrected wind speed and flux data from the recovered instrument data stream. * The telemetered FDCHP data contains an onboard processing error that makes the motion-corrected wind speed and flux data unusable. The instrument subject matter expert is being consulted, and the algorithms are under review. Only the processed, recovered data should be used. For further questions please email help@oceanobservatories.org.

Id: 1374 By: swhite
Flag: fail Exclude: No

CP01CNSM
12/13/14, 1:47 PM 12/15/14, 3:58 PM

Deployment 2: Short deployment due to the fuel cell not functioning correctly. No data expected.

Id: 709 By: cdobson

CP01CNSM-SBD12-08-FDCHPA000
Method: recovered_inst
Stream: fdchp_a_instrument_recovered
5/7/15, 1:34 PM

There are errors in the calculations of the recovered_instrument L1 and L2 data products (motion-corrected wind speeds, sonic temperature, and flux data products). This issue is under investigation (Redmine 13402).

Id: 1397 By: lgarzio

CP01CNSM-SBD12-08-FDCHPA000
10/23/15, 2:49 PM 4/4/16, 8:03 AM

Recovered data are missing.

Id: 544 By: leila

CP01CNSM-SBD12-08-FDCHPA000
10/23/15, 2:49 PM 5/12/16, 8:00 PM

Anemometer broken during deployment.

Id: 596 By: leila

CP01CNSM-SBD12-08-FDCHPA000
10/29/17, 10:15 AM 3/29/18, 3:37 PM

Deployment 8: Sonic temperature bugs: general slow response problem leading to errors of +/- 6 degrees from ambient, and an additional nonlinear error in the slow response when the ambient temperature is between -5 and 5 degrees. The fast response (> 1 Hz) of these sensors is still accurate. Therefore, flux measurements using the detrended temperature data should still be valid. w bug: positive vertical velocities w output by the anemometer need to be multiplied by 1.166; negative vertical velocities need to be multiplied by 1.289 (Redmine 11645)

Id: 2008 By: swhite
Flag: suspect Exclude: No

CP01CNSM-SBD12
4/23/18, 1:22 PM 10/29/18, 9:48 PM

Deployent 9: Power down - Instrument Disabled. No telemetered or recovered data expected for this time period.

Id: 1378 By: leila
Flag: not_operational Exclude: No

CP01CNSM-SBD12-08-FDCHPA000
10/29/18, 9:48 PM 4/7/19, 2:08 PM

Deployment 10: Sonic temperature bugs: general slow response problem leading to errors of +/- 6 degrees from ambient, and an additional nonlinear error in the slow response when the ambient temperature is between -5 and 5 degrees. The fast response (> 1 Hz) of these sensors is still accurate. Therefore, flux measurements using the detrended temperature data should still be valid. w bug: positive vertical velocities w output by the anemometer need to be multiplied by 1.166; negative vertical velocities need to be multiplied by 1.289 (Redmine 11645)

Id: 2009 By: swhite
Flag: suspect Exclude: No

CP01CNSM-SBD12-08-FDCHPA000
4/25/19, 8:00 PM 9/26/19, 1:15 PM

Deployment 11: Analysis of recovered instrument data shows that this instrument failed to initiaize beginning 2019-04-26. No telmetered or recovered data are available after this date for this deployment.

Id: 2010 By: cdobson
Flag: not_operational Exclude: No