Method | Data Stream | Content | Type | |
---|---|---|---|---|
recovered_wfp | parad_k__stc_imodem_instrument_recovered | Data Products | Report M2M Stats | Science |
telemetered | parad_k__stc_imodem_instrument | Data Products | Report M2M Stats | Science |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP02PMCI-WFP01-05-PARADK000 |
|||
CP02PMCI-WFP01-05-PARADK000 |
|||
CP02PMCI-WFP01 |
12/15/15, 12:00 AM | 5/20/16, 7:14 PM | |
CP02PMCI-WFP01-05-PARADK000 Deployment: 2 |
According to annotation ID 401, recovered data are missing. These data should be ingested or the annotation should be updated to indicate why the recovered data aren't available. By Lori Garzio, on 6/24/19 |
||
CP02PMCI-WFP01-05-PARADK000 Deployment: 5 |
5/21/16, 12:18 AM | 5/30/16, 12:05 PM |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP02PMCI-WFP01-05-PARADK000 Method: recovered_wfp |
10/11/14, 2:11 PM | 12/14/14, 7:00 PM |
Missing.
Id: 401
By: leila
|
CP02PMCI-WFP01 |
12/15/15, 10:08 PM | 1/29/16, 7:00 PM |
The Wire-Following Profiler was intentionally converted into a fixed-depth instrument to prolong the battery endurance. Id: 230 By: leila |
CP02PMCI |
5/20/16, 8:18 PM | 5/30/16, 8:05 AM |
Recovered data are not available because the Wire-Following Profiler motor and PAR cables were swapped and the data are thus unreliable. Id: 228 By: leila |
CP02PMCI |
10/2/16, 1:30 PM | 10/2/16, 2:30 PM |
Telemetered data are sparse because telemetry schedule was irregular during this deployment. Id: 232 By: leila |
CP02PMCI-WFP01 |
2/7/17, 7:00 AM | 6/16/17, 11:00 PM |
Data transmission interrupted. Id: 605 By: leila |
CP02PMCI-WFP01 |
12/31/19, 7:00 PM | 6/12/20, 12:01 PM |
Deployment 13: Due to a clock rollover issue, all WFPs have ceased sampling on 1/1/2020. No WFP data are being collected after this date. A resolution is being researched by the vendor. Id: 1881 By: cdobson |