Method | Data Stream | Content | Type | |
---|---|---|---|---|
recovered_wfp | dofst_k_wfp_instrument_recovered | Data Products | Report M2M Stats | Science |
recovered_wfp | dofst_k_wfp_metadata_recovered | Metadata | Report M2M | Engineering |
telemetered | dofst_k_wfp_instrument | Data Products | Report M2M Stats | Science |
telemetered | dofst_k_wfp_metadata | Metadata | Report M2M | Engineering |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP02PMUO-WFP01 |
11/23/13, 3:50 PM | 11/25/13, 12:00 AM | |
CP02PMUO-WFP01 |
2/3/15, 12:00 AM | 4/30/15, 1:00 AM | |
CP02PMUO-WFP01 |
8/20/15, 12:00 AM | 10/13/15, 1:04 PM |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP02PMUO-WFP01 |
11/23/13, 11:00 AM | 11/24/13, 3:28 PM |
Only few records available. The Wire-Following Profiler failed after few days and had to be recovered. Id: 732 By: leila |
CP02PMUO-WFP01 |
11/24/13, 3:28 PM | 11/24/13, 7:00 PM |
The Wire-Following Profiler was not performing correctly because of visible damage observed when recovered.
Id: 239
By: leila
|
CP02PMUO-WFP01 |
11/25/13, 3:00 PM | 4/14/14, 5:31 PM |
All communication to the Wire-Following Profiler were lost and no data are expected.
Id: 678
By: leila
|
CP02PMUO-WFP01 |
1/27/15, 4:00 AM | 4/29/15, 8:00 PM |
Telemetered data end on 27 Jan 2015. Id: 733 By: leila |
CP02PMUO-WFP01 |
2/3/15, 7:34 AM | 2/12/15, 1:00 AM |
Surface buoy mal functioning. Id: 735 By: leila |
CP02PMUO-WFP01 |
2/12/15, 4:12 PM | 4/29/15, 8:00 PM |
The surface buoy departed and had to be recovered. Id: 736 By: leila |
CP02PMUO-WFP01 |
10/2/15, 2:00 AM | 10/13/15, 9:04 AM |
Telemetered data end on 2 Oct 2015. Id: 734 By: leila |
CP02PMUO-WFP01 |
10/13/15, 3:12 PM | 5/18/16, 8:20 AM |
The Wire-Following Profiler failed.
Id: 428
By: leila
|
CP02PMUO-WFP01 |
8/18/16, 5:11 AM | 8/23/16, 5:00 AM |
The Wire-Following Profiler is stuck at the top of its range. The problem is assumed to be related to fouling of the line. Id: 43 |
CP02PMUO-WFP01 |
9/6/16, 5:13 AM | 10/1/16, 5:05 AM |
The Wire-Following Profiler is stuck at the top of its range. The problem is assumed to be related to fouling of the line. Id: 44 |
CP02PMUO-WFP01 |
10/23/16, 8:06 AM | 6/7/17, 5:05 AM |
The Wire-Following Profiler is stuck at the top of its range. The problem is assumed to be related to fouling of the line. Id: 45 By: leila |
CP02PMUO-WFP01 |
1/5/18, 10:00 AM | 4/10/18, 6:15 AM |
Deployment 10: The Wire-Following Profiler is stuck at the top of its range. Id: 939 By: leila |
CP02PMUO-WFP01 |
6/17/19, 8:00 PM | 10/14/19, 8:01 AM |
Deployment 13: WFP is likely stuck and profiles became erratic starting 2019-06-18. Id: 1626 By: cdobson |
CP02PMUO-WFP01 |
12/31/19, 7:00 PM | 6/9/20, 12:09 PM |
Deployment 14: Due to a clock rollover issue, all WFPs have ceased sampling on 1/1/2020. No WFP data are being collected after this date. A resolution is being researched by the vendor. Id: 1884 By: cdobson |