Asset: CGINS-FLORTK-01117

Type
Sensor
Mobile
0
Description of Equipment
3-Wavelength Fluorometer: FLORT Series K
Manufacturer
WET Labs
Model
ECO Triplet
Manufacturer Serial No
1117
Firmware Version
Acquisition Date
1/28/14
Original Cost
$0.00
Comments
Deployment Cruise Reference Designator Deployment Number Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
KN217 CP02PMCO-WFP01-04-FLORTK000 1 04/13/2014 10/07/2014 CGMCP-02PMCO-00001 CGVEH-CWFPCP-10301 CGINS-FLORTK-01117 40.0961 -70.8797 126 147
AT31 CP02PMCI-WFP01-04-FLORTK000 4 10/15/2015 05/20/2016 CGMCP-02PMCI-00004 CGVEH-CWFPCP-10301 CGINS-FLORTK-01117 40.2267 -70.8779 104 127
AR8 CP04OSPM-WFP01-04-FLORTK000 6 10/01/2016 06/01/2017 CGMCP-04OSPM-00006 CGVEH-CWFPCP-10301 CGINS-FLORTK-01117 39.9365 -70.8802 412 458
AR24 CP02PMUO-WFP01-04-FLORTK000 10 11/11/2017 04/10/2018 CGMCP-02PMUO-00010 CGVEH-CWFPCP-10301 CGINS-FLORTK-01117 39.9387 -70.7701 423 447
AR31 CP02PMUO-WFP01-04-FLORTK000 12 11/07/2018 04/22/2019 CGMCP-02PMUO-00012 CGVEH-CWFPCP-10301 CGINS-FLORTK-01117 39.9416 -70.7796 427 451
AR39 CP02PMUO-WFP01-04-FLORTK000 14 10/14/2019 06/09/2020 CGMCP-02PMUO-00014 CGVEH-CWFPCP-10301 CGINS-FLORTK-01117 39.9419 -70.7796 425 449

No notes yet.

Class Asset UID Start Date Serial Name Value Notes
FLORTK CGINS-FLORTK-01117 09/27/2013 1117 CC_angular_resolution 1.076
FLORTK CGINS-FLORTK-01117 09/27/2013 1117 CC_dark_counts_cdom 55
FLORTK CGINS-FLORTK-01117 09/27/2013 1117 CC_dark_counts_chlorophyll_a 50
FLORTK CGINS-FLORTK-01117 09/27/2013 1117 CC_dark_counts_volume_scatter 49
FLORTK CGINS-FLORTK-01117 09/27/2013 1117 CC_depolarization_ratio 0.039
FLORTK CGINS-FLORTK-01117 09/27/2013 1117 CC_measurement_wavelength 700.0
FLORTK CGINS-FLORTK-01117 09/27/2013 1117 CC_scale_factor_cdom 0.0916
FLORTK CGINS-FLORTK-01117 09/27/2013 1117 CC_scale_factor_chlorophyll_a 0.0126
FLORTK CGINS-FLORTK-01117 09/27/2013 1117 CC_scale_factor_volume_scatter 3.084e-06
FLORTK CGINS-FLORTK-01117 09/27/2013 1117 CC_scattering_angle 124.0
FLORTK CGINS-FLORTK-01117 02/12/2015 1117 CC_angular_resolution 1.076
FLORTK CGINS-FLORTK-01117 02/12/2015 1117 CC_dark_counts_cdom 48.0
FLORTK CGINS-FLORTK-01117 02/12/2015 1117 CC_dark_counts_chlorophyll_a 48.0
FLORTK CGINS-FLORTK-01117 02/12/2015 1117 CC_dark_counts_volume_scatter 50.0
FLORTK CGINS-FLORTK-01117 02/12/2015 1117 CC_depolarization_ratio 0.039
FLORTK CGINS-FLORTK-01117 02/12/2015 1117 CC_measurement_wavelength 700.0
FLORTK CGINS-FLORTK-01117 02/12/2015 1117 CC_scale_factor_cdom 0.0883
FLORTK CGINS-FLORTK-01117 02/12/2015 1117 CC_scale_factor_chlorophyll_a 0.0121
FLORTK CGINS-FLORTK-01117 02/12/2015 1117 CC_scale_factor_volume_scatter 2.913e-06
FLORTK CGINS-FLORTK-01117 02/12/2015 1117 CC_scattering_angle 124.0
FLORTK CGINS-FLORTK-01117 07/12/2016 1117 CC_angular_resolution 1.076
FLORTK CGINS-FLORTK-01117 07/12/2016 1117 CC_dark_counts_cdom 47.0
FLORTK CGINS-FLORTK-01117 07/12/2016 1117 CC_dark_counts_chlorophyll_a 47.0
FLORTK CGINS-FLORTK-01117 07/12/2016 1117 CC_dark_counts_volume_scatter 49.0
FLORTK CGINS-FLORTK-01117 07/12/2016 1117 CC_depolarization_ratio 0.039
FLORTK CGINS-FLORTK-01117 07/12/2016 1117 CC_measurement_wavelength 700.0
FLORTK CGINS-FLORTK-01117 07/12/2016 1117 CC_scale_factor_cdom 0.0713
FLORTK CGINS-FLORTK-01117 07/12/2016 1117 CC_scale_factor_chlorophyll_a 0.0118
FLORTK CGINS-FLORTK-01117 07/12/2016 1117 CC_scale_factor_volume_scatter 2.956e-06
FLORTK CGINS-FLORTK-01117 07/12/2016 1117 CC_scattering_angle 124.0
FLORTK CGINS-FLORTK-01117 07/25/2017 1117 CC_angular_resolution 1.076
FLORTK CGINS-FLORTK-01117 07/25/2017 1117 CC_dark_counts_cdom 50
FLORTK CGINS-FLORTK-01117 07/25/2017 1117 CC_dark_counts_chlorophyll_a 50
FLORTK CGINS-FLORTK-01117 07/25/2017 1117 CC_dark_counts_volume_scatter 49
FLORTK CGINS-FLORTK-01117 07/25/2017 1117 CC_depolarization_ratio 0.039
FLORTK CGINS-FLORTK-01117 07/25/2017 1117 CC_measurement_wavelength 700
FLORTK CGINS-FLORTK-01117 07/25/2017 1117 CC_scale_factor_cdom 0.0702
FLORTK CGINS-FLORTK-01117 07/25/2017 1117 CC_scale_factor_chlorophyll_a 0.0116
FLORTK CGINS-FLORTK-01117 07/25/2017 1117 CC_scale_factor_volume_scatter 3.159E-06
FLORTK CGINS-FLORTK-01117 07/25/2017 1117 CC_scattering_angle 124
FLORTK CGINS-FLORTK-01117 06/28/2018 1117 CC_angular_resolution 1.076
FLORTK CGINS-FLORTK-01117 06/28/2018 1117 CC_dark_counts_cdom 50
FLORTK CGINS-FLORTK-01117 06/28/2018 1117 CC_dark_counts_chlorophyll_a 50
FLORTK CGINS-FLORTK-01117 06/28/2018 1117 CC_dark_counts_volume_scatter 49
FLORTK CGINS-FLORTK-01117 06/28/2018 1117 CC_depolarization_ratio 0.039
FLORTK CGINS-FLORTK-01117 06/28/2018 1117 CC_measurement_wavelength 700
FLORTK CGINS-FLORTK-01117 06/28/2018 1117 CC_scale_factor_cdom 0.0622
FLORTK CGINS-FLORTK-01117 06/28/2018 1117 CC_scale_factor_chlorophyll_a 0.0114
FLORTK CGINS-FLORTK-01117 06/28/2018 1117 CC_scale_factor_volume_scatter 3.197E-06
FLORTK CGINS-FLORTK-01117 06/28/2018 1117 CC_scattering_angle 124
FLORTK CGINS-FLORTK-01117 06/25/2019 1117 CC_angular_resolution 1.076 source_file FLORT-K_BBFL2_SN_1117_Calibration_2019-06-25.zip > BBFL2-1117_CharSheet.pdf
FLORTK CGINS-FLORTK-01117 06/25/2019 1117 CC_dark_counts_cdom 47
FLORTK CGINS-FLORTK-01117 06/25/2019 1117 CC_dark_counts_chlorophyll_a 49
FLORTK CGINS-FLORTK-01117 06/25/2019 1117 CC_dark_counts_volume_scatter 49
FLORTK CGINS-FLORTK-01117 06/25/2019 1117 CC_depolarization_ratio 0.039
FLORTK CGINS-FLORTK-01117 06/25/2019 1117 CC_measurement_wavelength 700
FLORTK CGINS-FLORTK-01117 06/25/2019 1117 CC_scale_factor_cdom 0.0553
FLORTK CGINS-FLORTK-01117 06/25/2019 1117 CC_scale_factor_chlorophyll_a 0.0114
FLORTK CGINS-FLORTK-01117 06/25/2019 1117 CC_scale_factor_volume_scatter 3.240E-06
FLORTK CGINS-FLORTK-01117 06/25/2019 1117 CC_scattering_angle 124