3-Wavelength Fluorometer

Reference Designator
CP02PMUO-WFP01-04-FLORTK000
Start Depth
15
End Depth
425
Location
Wire Following Profiler Body
Current Status
Review Complete
Class
FLORT (3-Wavelength Fluorometer)
Series
FLORT-K
Science Discipline
Biological
Make
WET Labs
Model
ECO Triplet
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CP02PMUO/WFP01/04-FLORTK000/metadata
Method Data Stream Content Type
recovered_wfp flort_sample Data Products Report M2M Stats Science
telemetered flort_sample Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review KN214 11/23/2013 11/25/2013 CGMCP-02PMUO-00001 CGVEH-CWFPCP-93601 CGINS-FLORTK-00969 39.9416 -70.7797 421 445
2 Review KN214 11/25/2013 04/14/2014 CGMCP-02PMUO-00002 CGVEH-CWFPCP-99102 CGINS-FLORTK-01031 39.938 -70.7691 426 450
3 Review KN217 04/14/2014 10/05/2014 CGMCP-02PMUO-00003 CGVEH-CWFPCP-10304 CGINS-FLORTK-01100 39.9424 -70.7801 428 452
4 Review KN222 10/05/2014 04/30/2015 CGMCP-02PMUO-00004 CGVEH-CWFPCP-99105 CGINS-FLORTK-01033 39.943 -70.77 426 450
5 Review AT27 04/30/2015 10/13/2015 CGMCP-02PMUO-00005 CGVEH-CWFPCP-10304 CGINS-FLORTK-01100 39.9416 -70.7809 428 452
6 Review AT31 10/13/2015 05/18/2016 CGMCP-02PMUO-00006 CGVEH-CWFPCP-99105 CGINS-FLORTK-01033 39.9434 -70.7704 420 444
7 Review AR4 05/18/2016 10/01/2016 CGMCP-02PMUO-00007 CGVEH-CWFPCP-10304 CGINS-FLORTK-01100 39.9421 -70.78 426 450
8 Review AR8 10/02/2016 06/07/2017 CGMCP-02PMUO-00008 CGVEH-CWFPCP-99105 CGINS-FLORTK-01466 39.9394 -70.7708 423 447
9 Review AR18 06/07/2017 11/11/2017 CGMCP-02PMUO-00009 CGVEH-CWFPCP-10304 CGINS-FLORTK-01100 39.9412 -70.7804 425 449
10 Review AR24 11/11/2017 04/10/2018 CGMCP-02PMUO-00010 CGVEH-CWFPCP-10301 CGINS-FLORTK-01117 39.9387 -70.7701 423 447
11 Review AR28 04/09/2018 11/07/2018 CGMCP-02PMUO-00011 CGVEH-CWFPCP-10304 CGINS-FLORTK-01100 39.9419 -70.779 426 450
12 Review AR31 11/07/2018 04/22/2019 CGMCP-02PMUO-00012 CGVEH-CWFPCP-10301 CGINS-FLORTK-01117 39.9416 -70.7796 427 451
13 Review AR34 04/23/2019 10/14/2019 CGMCP-02PMUO-00013 CGVEH-CWFPCP-10304 CGINS-FLORTK-01100 39.9417 -70.7804 425 449
14 Review AR39 10/14/2019 06/09/2020 CGMCP-02PMUO-00014 CGVEH-CWFPCP-10301 CGINS-FLORTK-01117 39.9419 -70.7796 425 449
15 Review AR44 06/09/2020 CGMCP-02PMUO-00015 CGVEH-CWFPCP-10304 CGINS-FLORTK-01100 39.9429 -70.7709 419 443
Metadata Start Date End Date Comment
CP02PMUO-WFP01-04-FLORTK000

The pressure coordinate (pressure_depth) is an array of all fill values. This variable should contain valid pressure data. There is another variable in the files called in_ctd_pressure that contains what appears to be valid pressure data. This is the case for every FLORTK on every WFP. https://opendap.oceanobservatories.org/thredds/dodsC/ooi/lgarzio@marine.rutgers.edu/20181218T153703-CP02PMUO-WFP01-04-FLORTK000-recovered_wfp-flort_sample/deployment0010_CP02PMUO-WFP01-04-FLORTK000-recovered_wfp-flort_sample_20171112T050202-20180410T090344.nc.ascii?pressure_depth[0:1:1000],int_ctd_pressure[0:1:1000]

By Lori Garzio, on 3/12/19

CP02PMUO-WFP01-04-FLORTK000

Comparisons with the shipboard CTD casts were done for D0003, R0003, D0004, D0005, R0005, D0007, R0007, D0008, R0008, D0009, R0009. Chl-a recorded by the profiler was slightly higher at deployment compared to the CTD cast, particularly at the chl-a maximum (with the exception of D0005 where the profiler was lower), and substantially higher compared to the CTD cast at recovery. Profiler values for R0007 are unreasonable. Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/13/19

CP02PMUO-WFP01
11/23/13, 3:50 PM 11/25/13, 12:00 AM

Deployment 1: suspect data - see annotation IDs 239 and 732

By Lori Garzio, on 8/23/19

CP02PMUO-WFP01
2/3/15, 12:00 AM 4/30/15, 1:00 AM

Deployment 4: surface buoy malfunctioned and then detached from the mooring at the end of the deployment (annotation IDs 735 and 736) - suspect data.

By Lori Garzio, on 8/23/19

CP02PMUO-WFP01
8/20/15, 12:00 AM 10/13/15, 1:04 PM

Deployment 5: Suspect data

By Lori Garzio, on 8/23/19

CP02PMUO-WFP01-04-FLORTK000
Deployment: 3
7/20/14, 12:00 AM 10/5/14, 12:11 PM

Suspect data towards the end of the deployment.

By Lori Garzio, on 8/23/19

CP02PMUO-WFP01-04-FLORTK000
Deployment: 4

71% of CDOM values for this deployment are outside of global ranges

By Lori Garzio, on 8/23/19

CP02PMUO-WFP01-04-FLORTK000
Deployment: 7
8/20/16, 12:00 AM 10/1/16, 4:16 PM

Suspect data towards the end of the deployment

By Lori Garzio, on 8/23/19

CP02PMUO-WFP01-04-FLORTK000
Deployment: 8

63% of total_volume_scattering_coefficient values are outside of global ranges for this deployment

By Lori Garzio, on 8/23/19

CP02PMUO-WFP01-04-FLORTK000
Deployment: 8
10/25/16, 12:00 AM 6/7/17, 12:08 PM

Suspect backscatter

By Lori Garzio, on 8/23/19

CP02PMUO-WFP01-04-FLORTK000
Deployment: 9
8/28/17, 12:00 AM 11/11/17, 12:45 PM

Suspect data at the end of the deployment

By Lori Garzio, on 8/23/19

CP02PMUO-WFP01-04-FLORTK000
Deployment: 10
2/15/18, 12:00 AM 4/10/18, 10:15 AM

Suspect data at the end of the deployment

By Lori Garzio, on 8/23/19

New Note

Metadata Start Date End Date Comment
CP02PMUO-WFP01
11/23/13, 11:00 AM 11/24/13, 3:28 PM

Only few records available. The Wire-Following Profiler failed after few days and had to be recovered.

Id: 732 By: leila

CP02PMUO-WFP01
11/24/13, 3:28 PM 11/24/13, 7:00 PM

The Wire-Following Profiler was not performing correctly because of visible damage observed when recovered.

Id: 239 By: leila
Flag: not_operational Exclude: No

CP02PMUO-WFP01
11/25/13, 3:00 PM 4/14/14, 5:31 PM

All communication to the Wire-Following Profiler were lost and no data are expected.

Id: 678 By: leila
Flag: not_operational Exclude: No

CP02PMUO-WFP01
1/27/15, 4:00 AM 4/29/15, 8:00 PM

Telemetered data end on 27 Jan 2015.

Id: 733 By: leila

CP02PMUO-WFP01
2/3/15, 7:34 AM 2/12/15, 1:00 AM

Surface buoy mal functioning.

Id: 735 By: leila

CP02PMUO-WFP01
2/12/15, 4:12 PM 4/29/15, 8:00 PM

The surface buoy departed and had to be recovered.

Id: 736 By: leila

CP02PMUO-WFP01
10/2/15, 2:00 AM 10/13/15, 9:04 AM

Telemetered data end on 2 Oct 2015.

Id: 734 By: leila

CP02PMUO-WFP01
10/13/15, 3:12 PM 5/18/16, 8:20 AM

The Wire-Following Profiler failed.

Id: 428 By: leila
Flag: not_available Exclude: No

CP02PMUO-WFP01
8/18/16, 5:11 AM 8/23/16, 5:00 AM

The Wire-Following Profiler is stuck at the top of its range. The problem is assumed to be related to fouling of the line.

Id: 43

CP02PMUO-WFP01
9/6/16, 5:13 AM 10/1/16, 5:05 AM

The Wire-Following Profiler is stuck at the top of its range. The problem is assumed to be related to fouling of the line.

Id: 44

CP02PMUO-WFP01
10/23/16, 8:06 AM 6/7/17, 5:05 AM

The Wire-Following Profiler is stuck at the top of its range. The problem is assumed to be related to fouling of the line.

Id: 45 By: leila

CP02PMUO-WFP01
1/5/18, 10:00 AM 4/10/18, 6:15 AM

Deployment 10: The Wire-Following Profiler is stuck at the top of its range.

Id: 939 By: leila

CP02PMUO-WFP01
6/17/19, 8:00 PM 10/14/19, 8:01 AM

Deployment 13: WFP is likely stuck and profiles became erratic starting 2019-06-18.

Id: 1626 By: cdobson

CP02PMUO-WFP01
12/31/19, 7:00 PM 6/9/20, 12:09 PM

Deployment 14: Due to a clock rollover issue, all WFPs have ceased sampling on 1/1/2020. No WFP data are being collected after this date. A resolution is being researched by the vendor.

Id: 1884 By: cdobson