3-Wavelength Fluorometer

Reference Designator
CP02PMCI-WFP01-04-FLORTK000
Review Status
Review Complete
Note
Depth
15 to 104m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Triplet

Dataset Reviews Last processed: 6/17/19, 5:13 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_wfp flort_sample 178 177 0 0 0 0 170,603 5 104 / Complete
2 telemetered flort_sample 66 49 0 16 0 0 43,277 5 104 / 1 1 Complete
3 recovered_wfp flort_sample 169 168 1 0 0 0 130,105 6 104 / Complete
4 recovered_wfp flort_sample 219 219 0 0 0 0 459,601 5 104 / Complete
5 telemetered flort_sample 10 10 0 0 0 0 11,959 5 104 / 1 1 Complete
6 recovered_wfp flort_sample 94 93 0 0 0 0 96,323 5 104 / Complete
7 recovered_wfp flort_sample 258 257 0 0 0 0 237,975 5 104 / Complete
8 recovered_wfp flort_sample 145 132 0 11 0 0 128,503 5 104 / Complete
9 recovered_wfp flort_sample 156 156 0 0 0 0 118,541 6 104 / Complete
Data Ranges Review Images

Test Notes

  1. no other streams for comparison

Data Coverage

Deployment: 123456789
flort_sample 99%74%99%100%100%99%100%91%100%

Lat/Lon Differences (km)

Deployment: 123456789
1 0.00
2 0.900.00
3 0.020.920.00
4 0.910.020.930.00
5 0.040.940.020.940.00
6 0.070.970.060.980.050.00
7 0.890.030.900.030.920.950.00
8 0.040.930.030.940.050.040.910.00
9 0.890.020.900.030.920.950.010.910.00

System Annotations

Metadata Start Date End Date Comment
CP02PMCI-WFP01-04-FLORTK000
Method: recovered_wfp
10/11/14, 2:11 PM 12/14/14, 7:00 PM

Missing.

Id: 400 By: leila
Flag: not_available Exclude: No

CP02PMCI-WFP01
12/15/15, 10:08 PM 1/29/16, 7:00 PM

The Wire-Following Profiler was intentionally converted into a fixed-depth instrument to prolong the battery endurance.

Id: 230 By: leila

CP02PMCI
5/20/16, 8:18 PM 5/30/16, 8:05 AM

Recovered data are not available because the Wire-Following Profiler motor and PAR cables were swapped and the data are thus unreliable.

Id: 228 By: leila

CP02PMCI
10/2/16, 1:30 PM 10/2/16, 2:30 PM

Telemetered data are sparse because telemetry schedule was irregular during this deployment.

Id: 232 By: leila

CP02PMCI-WFP01
2/7/17, 7:00 AM 6/16/17, 11:00 PM

Data transmission interrupted.

Id: 605 By: leila

CP02PMCI-WFP01
12/31/19, 7:00 PM 6/12/20, 12:01 PM

Deployment 13: Due to a clock rollover issue, all WFPs have ceased sampling on 1/1/2020. No WFP data are being collected after this date. A resolution is being researched by the vendor.

Id: 1881 By: cdobson

Review Notes

Metadata Start Date End Date Comment
CP02PMCI-WFP01-04-FLORTK000

The pressure coordinate (pressure_depth) is an array of all fill values. This variable should contain valid pressure data. There is another variable in the files called in_ctd_pressure that contains what appears to be valid pressure data. This is the case for every FLORTK on every WFP. https://opendap.oceanobservatories.org/thredds/dodsC/ooi/lgarzio@marine.rutgers.edu/20181212T200946-CP02PMCI-WFP01-04-FLORTK000-recovered_wfp-flort_sample/deployment0009_CP02PMCI-WFP01-04-FLORTK000-recovered_wfp-flort_sample_20171107T230203-20180411T133621.nc.ascii?pressure_depth[0:1:1000],int_ctd_pressure[0:1:1000]

By Lori Garzio, on 3/12/19

CP02PMCI-WFP01-04-FLORTK000

Comparisons with the shipboard CTD casts were done for D0001, R0001, D0003, R0003, D0004, R0004, D0006, R0006, D0007, R0007, D0008, D0009, R0009. Chl-a recorded by the profiler was slightly higher at deployment compared to the CTD cast (with the exception of D0003), and substantially higher compared to the CTD cast at recovery. Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/13/19

CP02PMCI-WFP01-04-FLORTK000

It looks like biofouling might be an issue for this instrument. Backscatter, CDOM, and chl-a values are unreasonable during the second half of almost every deployment.

By Lori Garzio, on 6/24/19

CP02PMCI-WFP01
12/15/15, 12:00 AM 5/20/16, 7:14 PM

Annotation ID 230: Deployment 4: WFP stopped profiling. Excluding from final data ranges.

By Lori Garzio, on 6/24/19

CP02PMCI-WFP01-04-FLORTK000
Deployment: 1
7/10/14, 12:00 AM 10/10/14, 10:38 AM

Suspect data for all FLORT variables - this should be investigated and annotated.

By Lori Garzio, on 6/24/19

CP02PMCI-WFP01-04-FLORTK000
Deployment: 2

According to annotation ID 400, recovered data are missing. These data should be ingested or the annotation should be updated to indicate why the recovered data aren't available.

By Lori Garzio, on 6/24/19

CP02PMCI-WFP01-04-FLORTK000
Deployment: 3
7/29/15, 12:00 AM 10/15/15, 12:03 PM

Suspect data for all FLORT variables - this should be investigated and annotated.

By Lori Garzio, on 6/24/19

CP02PMCI-WFP01-04-FLORTK000
Deployment: 6
8/20/16, 12:00 AM 10/2/16, 8:32 PM

Suspect data for all FLORT variables - this should be investigated and annotated.

By Lori Garzio, on 6/24/19

CP02PMCI-WFP01-04-FLORTK000
Deployment: 7
11/15/16, 12:00 AM 6/17/17, 12:12 PM

Suspect data for all FLORT variables - this should be investigated and annotated.

By Lori Garzio, on 6/24/19

CP02PMCI-WFP01-04-FLORTK000
Deployment: 8
8/15/17, 12:00 AM 11/7/17, 12:57 PM

Suspect data for all FLORT variables - this should be investigated and annotated.

By Lori Garzio, on 6/24/19

CP02PMCI-WFP01-04-FLORTK000
Deployment: 9
2/15/18, 12:00 AM 4/11/18, 1:46 PM

Suspect data for all FLORT variables.

By Lori Garzio, on 6/24/19