Method | Data Stream | Content | Type | |
---|---|---|---|---|
recovered_host | pco2a_a_dcl_instrument_air_recovered | In-Air Data Products | Report M2M Stats | Science |
recovered_host | pco2a_a_dcl_instrument_water_recovered | In-Water Data Products | Report M2M Stats | Science |
telemetered | pco2a_a_dcl_instrument_air | In-Air Data Products | Report M2M Stats | Science |
telemetered | pco2a_a_dcl_instrument_water | In-Water Data Products | Report M2M Stats | Science |
Deployment | Cruise | Start Date | Stop Date | Mooring Asset | Node Asset | Sensor Asset | Latitude | Longitude | Deployment Depth | Water Depth |
---|---|---|---|---|---|---|---|---|---|---|
1 Review | OC1503D | 04/07/2015 | 05/09/2016 | CGMCE-04OSSM-00003 | CGINS-PCO2AA-33199 | 44.3667 | -124.945 | 0.7 | 587 | |
2 Review | TN-342 | 05/16/2016 | 10/01/2016 | CGMCE-04OSSM-00002 | CGINS-PCO2AA-34198 | 44.3648 | -124.943 | 0.7 | 587 | |
3 Review | AT37-03 | 10/01/2016 | 04/21/2017 | CGMCE-04OSSM-00003 | CGINS-PCO2AA-35256 | 44.3811 | -124.956 | 0.7 | 585 | |
4 Review | SKQ201704S | 04/21/2017 | 10/13/2017 | CGMCE-04OSSM-00004 | CGINS-PCO2AA-34198 | 44.3656 | -124.941 | 0.7 | 574 | |
5 Review | SKQ201715S | 10/11/2017 | 04/03/2018 | CGMCE-04OSSM-00005 | CGINS-PCO2AA-35257 | 44.3818 | -124.95 | 0.7 | 575 | |
6 Review | SKQ201808S | 04/03/2018 | 09/27/2018 | CGMCE-04OSSM-00006 | CGINS-PCO2AA-34197 | 44.3661 | -124.94 | 0.7 | 575 | |
7 Review | SR1813 | 09/27/2018 | 04/20/2019 | CGMCE-04OSSM-00007 | CGINS-PCO2AA-35257 | 44.38 | -124.95 | 0.7 | 575 | |
8 Review | SKQ201910S | 04/20/2019 | 10/21/2019 | CGMCE-04OSSM-00008 | CGINS-PCO2AA-34230 | 44.3645 | -124.94 | 0.7 | 575 | |
9 Review | SKQ201921S | 10/21/2019 | CGMCE-04OSSM-00009 | CGINS-PCO2AA-35257 | 44.378 | -124.946 | 0.7 | 575 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CE04OSSM-SBD12-04-PCO2AA000 Deployment: 1 |
4/30/15, 12:00 AM | 5/2/15, 12:00 AM | |
CE04OSSM-SBD12-04-PCO2AA000 Deployment: 2 |
8/29/16, 12:00 AM | 10/1/16, 3:21 PM | |
CE04OSSM-SBD12-04-PCO2AA000 Deployment: 3 |
measured_air_co2 values at the end of deployment 3 (4/21/2017) are lower than the values recorded at the beginning of deployment 4. By Lori Garzio, on 7/29/19 |
||
CE04OSSM-SBD12-04-PCO2AA000 Deployment: 5 |
recovered_host data are not available for download for this deployment. According to the ingest csv, data aren't expected because the SD card was corrupted. This should be annotated in the system. By Lori Garzio, on 7/26/19 |
||
CE04OSSM-SBD12-04-PCO2AA000 Deployment: 6 |
4/3/18, 6:16 PM | 9/27/18, 7:35 PM |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CE04OSSM-SBD12-04-PCO2AA000 Method: telemetered Stream: pco2a_a_dcl_instrument_air |
4/30/15, 6:30 AM | 4/30/15, 10:30 PM |
A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.
Id: 1949
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_air_recovered |
4/30/15, 6:30 AM | 4/30/15, 10:30 PM |
A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.
Id: 1954
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: telemetered Stream: pco2a_a_dcl_instrument_water |
4/30/15, 6:30 AM | 4/30/15, 10:30 PM |
A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.
Id: 1960
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_water_recovered |
4/30/15, 6:30 AM | 4/30/15, 10:30 PM |
A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.
Id: 1964
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: telemetered Stream: pco2a_a_dcl_instrument_air |
9/10/16, 2:30 PM | 10/1/16, 5:30 AM |
A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.
Id: 1950
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_air_recovered |
9/10/16, 2:30 PM | 10/1/16, 5:30 AM |
A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.
Id: 1955
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: telemetered Stream: pco2a_a_dcl_instrument_water |
9/10/16, 2:30 PM | 10/1/16, 5:30 AM |
A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.
Id: 1961
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_water_recovered |
9/10/16, 2:30 PM | 10/1/16, 5:30 AM |
A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.
Id: 1965
By: wingardc
|
CE04OSSM-SBD12 Method: recovered_host |
10/11/17, 7:08 PM | 4/3/18, 4:27 PM |
The SD card with the recovered host data for deployment 5 was corrupted and we were unable to download this data set. Please use the telemetered data instead as that data is identical to data available via this method.
Id: 1823
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_air_recovered |
4/3/18, 2:30 PM | 9/27/18, 3:30 PM |
A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2018 deployment (deployment 06), nor the associated sea-air pCO2 flux estimate.
Id: 1956
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: telemetered Stream: pco2a_a_dcl_instrument_air |
4/3/18, 2:30 PM | 9/27/18, 3:30 PM |
UPDATED, 2020-01-28: The QC flag for the air measurement has been reset to fail. Users should not use the PCO2A air measurement for the Spring 2018 deployment (deployment 6), nor the associated sea-air pCO2 flux estimate. *PCO2A air measurement for the Spring 2018 deployment is considered suspect. Recent investigations have indicated a vendor error in assembly may have pinched closed an internal tube carrying the surface air stream into the sample volume resulting in an erroneous measurement.
Id: 1447
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: telemetered Stream: pco2a_a_dcl_instrument_water |
4/3/18, 2:30 PM | 9/27/18, 3:30 PM |
A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2018 deployment (deployment 06), nor the associated sea-air pCO2 flux estimate.
Id: 1990
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_water_recovered |
4/3/18, 2:30 PM | 9/27/18, 3:30 PM |
A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2018 deployment (deployment 06), nor the associated sea-air pCO2 flux estimate.
Id: 1991
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: telemetered Stream: pco2a_a_dcl_instrument_air |
12/8/18, 8:30 AM | 12/8/18, 11:30 PM |
A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.
Id: 1951
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_air_recovered |
12/8/18, 8:30 AM | 12/8/18, 11:30 PM |
A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.
Id: 1957
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: telemetered Stream: pco2a_a_dcl_instrument_air |
3/8/19, 5:30 PM | 4/21/19, 12:30 PM |
A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.
Id: 1952
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_air_recovered |
3/8/19, 5:30 PM | 4/21/19, 12:30 PM |
A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.
Id: 1958
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: telemetered Stream: pco2a_a_dcl_instrument_water |
3/8/19, 5:30 PM | 4/21/19, 12:30 PM |
A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.
Id: 1962
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_water_recovered |
3/8/19, 5:30 PM | 4/21/19, 12:30 PM |
A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.
Id: 1966
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: telemetered |
4/20/19, 4:13 PM | 8/15/19, 11:30 AM |
UPDATED, 2019-09-04: The vendor provided a protocol to reset the zero measurement on a daily basis. This protocol has been implemented on the mooring and the pCO2 air and water measurements have reset. They are still somewhat low, and should be considered as suspect. The vendor will work to provided a finalized and corrected dataset after the moorings are recovered and the sensors are returned for servicing. *Due to a firmware bug, the sensor zeroing function for correcting the baseline drift over time is not currently saving the updated reference value used to compute the raw CO2 value outputs during the measurements. The sensors are zeroing at the end of the sample cycle when scheduled, but the firmware is not storing the new value. This results in the original zero reference value being used for all measurements and an observable and consistent negative drift in both the air and water CO2 values. The vendor has indicated that the data may be correctable post-recovery, however at this time no such correction algorithm has been provided.
Id: 1682
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: telemetered |
10/28/19, 2:30 PM | 10/31/19, 2:30 PM |
The data logger port providing power to the instrument showed an over-current error and disabled the instrument. The port had to be power cycled in order to clear the error and restore operations.
Id: 2053
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: telemetered Stream: pco2a_a_dcl_instrument_air |
11/22/19, 12:30 PM | 11/23/19, 1:30 PM |
A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.
Id: 1953
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_air_recovered |
11/22/19, 12:30 PM | 11/23/19, 1:30 PM |
A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.
Id: 1959
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: telemetered Stream: pco2a_a_dcl_instrument_water |
11/22/19, 12:30 PM | 11/23/19, 1:30 PM |
A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.
Id: 1963
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_water_recovered |
11/22/19, 12:30 PM | 11/23/19, 1:30 PM |
A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.
Id: 1967
By: wingardc
|
CE04OSSM-SBD12-04-PCO2AA000 Method: telemetered |
2/23/20, 6:30 PM | 3/1/20, 1:30 PM |
The instrument suffered a firmware crash and had to be power cycled in order to restore operations. We are in consultation with the vendor to determine the cause of this failure.
Id: 2056
By: wingardc
|
CE04OSSM Method: telemetered |
5/6/20, 10:40 PM | 5/11/20, 8:00 PM |
Due to a bug in the mooring telemetry system, data during this time range was not telemetered to shore. It was recorded on the mooring and instruments and will be available in the recovered data streams.
Id: 2181
By: wingardc
|