Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry. By Lori Garzio, on 5/10/19 |
||
CE05MOAS |
Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of 3 consecutive down and up casts for CE05MOAS-GL312-04-DOSTAM000 on 7/8/2014 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue. By Lori Garzio, on 5/14/19 |
||
CE05MOAS-GL386 |
Short 2-3 day deployments for deployments 6-9 should be annotated. By Lori Garzio, on 8/30/19 |
||
CE05MOAS-GL386 Deployment: 3 |
|||
CE05MOAS-GL386 Deployment: 5 |
|||
CE05MOAS-GL386-02-FLORTM000 Deployment: 5 |
Values for every parameter are close to zero for the entire deployment. At recovery, black electrical tape was still covering the sensor (Redmine 12049). This should be annotated in uframe. By Lori Garzio, on 8/30/19 |
||
CE05MOAS-GL386-02-FLORTM000 Deployment: 7 |
The deployment 7 .nc files are missing coordinates lat, lon, and pressure. By Lori Garzio, on 8/30/19 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CE05MOAS-GL386-02-FLORTM000 |
3/1/17, 7:00 PM | 3/7/17, 7:00 PM |
Instrument was turned off on 2017-03-02 to prolong battery life to recovery date.
Id: 895
By: michaesm
|