Method | Data Stream | Content | Type | |
---|---|---|---|---|
recovered_wfp | ctdpf_ckl_wfp_instrument_recovered | Data Products | Report M2M Stats | Science |
recovered_wfp | ctdpf_ckl_wfp_metadata_recovered | Metadata | Report M2M | Engineering |
telemetered | ctdpf_ckl_wfp_instrument | Data Products | Report M2M Stats | Science |
telemetered | ctdpf_ckl_wfp_metadata | Metadata | Report M2M | Engineering |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP02PMCI-WFP01-03-CTDPFK000 |
Comparisons with the shipboard CTD casts were done for D0001, R0001, D0003, R0003, D0004, R0004, D0006, R0006, D0007, R0007, D0008, D0009, R0009. The shapes of the profiles generally look similar, however the values are often different (particularly at shallower depths). See attached plots for details. Comparison to the shipboard bottle data is outside the scope of this project. By Lori Garzio, on 3/13/19 |
||
CP02PMCI-WFP01 |
12/15/15, 12:00 AM | 5/20/16, 7:14 PM | |
CP02PMCI-WFP01-03-CTDPFK000 Deployment: 2 |
According to annotation ID 399, recovered data are missing. These data should be ingested or the annotation should be updated to indicate why the recovered data aren't available. By Lori Garzio, on 6/21/19 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP02PMCI-WFP01-03-CTDPFK000 Method: recovered_wfp |
10/11/14, 2:11 PM | 12/14/14, 7:00 PM |
Missing.
Id: 399
By: leila
|
CP02PMCI-WFP01 |
12/15/15, 10:08 PM | 1/29/16, 7:00 PM |
The Wire-Following Profiler was intentionally converted into a fixed-depth instrument to prolong the battery endurance. Id: 230 By: leila |
CP02PMCI |
5/20/16, 8:18 PM | 5/30/16, 8:05 AM |
Recovered data are not available because the Wire-Following Profiler motor and PAR cables were swapped and the data are thus unreliable. Id: 228 By: leila |
CP02PMCI |
10/2/16, 1:30 PM | 10/2/16, 2:30 PM |
Telemetered data are sparse because telemetry schedule was irregular during this deployment. Id: 232 By: leila |
CP02PMCI-WFP01 |
2/7/17, 7:00 AM | 6/16/17, 11:00 PM |
Data transmission interrupted. Id: 605 By: leila |
CP02PMCI-WFP01 |
12/31/19, 7:00 PM | 6/12/20, 12:01 PM |
Deployment 13: Due to a clock rollover issue, all WFPs have ceased sampling on 1/1/2020. No WFP data are being collected after this date. A resolution is being researched by the vendor. Id: 1881 By: cdobson |