Photosynthetically Available Radiation

Reference Designator
CP05MOAS-GL387-05-PARADM000
Start Depth
0
End Depth
1,000
Location
Current Status
Review Complete
Class
PARAD (Photosynthetically Available Radiation)
Series
PARAD-M
Science Discipline
Biological
Make
Biospherical Instruments
Model
QSP-2155
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CP05MOAS/GL387/05-PARADM000/metadata
Method Data Stream Content Type
recovered_host parad_m_glider_recovered Data Products Report M2M Stats Science
telemetered parad_m_glider_instrument Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review KN-217 04/16/2014 05/02/2014 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00387 CGINS-PARADM-50165 39.9617 -70.9058 1000 1000
2 Review SI20150110 01/11/2015 03/04/2015 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00387 CGINS-PARADM-50165 40.0098 -70.3986 1000 1000
3 Review AT-31 10/14/2015 11/30/2015 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00387 CGINS-PARADM-50165 40.07 -70.4152 1000 1000
4 Review AR1-07 04/04/2016 07/29/2016 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00387 CGINS-PARADM-50165 39.9685 -70.5384 1000 1000
5 Review SI20170418 04/19/2017 07/02/2017 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00387 CGINS-PARADM-50167 39.7836 -70.4657 1000 1000
6 Review AR-24 10/28/2017 01/03/2018 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00387 CGINS-PARADM-50167 40.0318 -70.6204 1000 1000
7 Review AR-34 04/08/2019 06/18/2019 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00387 CGINS-PARADM-50146 39.9681 -71 1000 1000
Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-GL387-05-PARADM000

99% of the sci_bsipar_par values for deployments 1-4 are outside of global ranges. sci_bsipar_par is the unscaled L0 PAR data product. The global ranges for this parameter are the same as for the scaled L1 PAR data product [0.0, 2500.0]. The global range for sci_bsipar_par should be reviewed.

By Lori Garzio, on 9/5/19

CP05MOAS-GL387
Deployment: 1

Glider doesn't look like it sampled according to plan. This should be annotated.

By Lori Garzio, on 9/5/19

CP05MOAS-GL387-05-PARADM000
Deployment: 5

parad_m_par values are orders of magnitude too low for this deployment. sci_bsipar_par values are reasonable, so the scaling factor should be reviewed.

By Lori Garzio, on 9/5/19

CP05MOAS-GL387-05-PARADM000
Deployment: 6

parad_m_par values are orders of magnitude too low for this deployment. sci_bsipar_par values are reasonable, so the scaling factor should be reviewed.

By Lori Garzio, on 9/5/19

CP05MOAS-GL387-05-PARADM000
Deployment: 6

50% of the sci_bsipar_par and parad_m_par data for this deployment are outside of global ranges [0.0, 2500.0]. The majority of the values outside of global ranges are slightly negative

By Lori Garzio, on 9/5/19

New Note

Metadata Start Date End Date Comment
CP05MOAS-GL387
7/13/16, 9:00 PM 7/28/16, 8:00 AM

The glider altimeter may have failed.

Id: 644 By: leila
Flag: not_available Exclude: No

CP05MOAS-GL387
5/26/19, 8:00 PM

Deployment 7: Glider's altimeter failed on 5/27/2019 so it will be limited to sampling the bottom line of the FZ box and to 500m until the end of the deployment.

Id: 1620 By: cdobson