3-Wavelength Fluorometer

Reference Designator
GS01SUMO-SBD12-02-FLORTD000
Start Depth
1
End Depth
1
Location
Bottom of Buoy
Current Status
Review Complete
Class
FLORT (3-Wavelength Fluorometer)
Series
FLORT-D
Science Discipline
Biological
Make
WET Labs
Model
ECO Triplet-w
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/GS01SUMO/SBD12/02-FLORTD000/metadata
Method Data Stream Content Type
recovered_host flort_sample Data Products Report M2M Stats Science
telemetered flort_sample Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review AT-26-29 02/18/2015 12/27/2015 CGMGS-01SUMO-00001 CGINS-FLORTD-01212 -54.4082 -89.3576 1 4611
2 Review NBP-15-11 12/14/2015 12/12/2016 CGMGS-01SUMO-00002 CGINS-FLORTD-01319 -54.4041 -89.2069 1 4588
3 Review NBP16-10 11/25/2016 12/09/2018 CGMGS-01SUMO-00003 CGINS-FLORTD-01218 -54.4076 -89.3567 1 4612
4 Review DY096 12/04/2018 01/20/2020 CGMGS-01SUMO-00004 CGINS-FLORTD-01107 -54.4072 -89.206 1 4589
Metadata Start Date End Date Comment
GS01SUMO-SBD12-02-FLORTD000

pressure_depth (dbar) is an array of fill values in all files. It's unclear if this variable should contain valid data, or if it's assumed that the pressure is 1 because this instrument is on the surface buoy.

By Lori Garzio, on 3/22/19

GS01SUMO-SBD12-02-FLORTD000
Deployment: 2
7/1/16, 12:00 AM 12/12/16, 7:58 AM

chl-a concentrations towards the end of the deployment are unreasonable, possibly from biofouling? Issue should be investigated and annotated. Excluding from data range summary.

By Lori Garzio, on 3/22/19

GS01SUMO
Deployment: 3

Annotation ID 834 needs to be updated: "Deployment 3: Recovery of the mooring was planned during the NBP 17-09 cruise (Nov - Dec 2017) but not completed due to adverse weather conditions. Recovery options are pending. Instruments that are still operational are likely reaching (or have surpassed) their deployment lifespan."

By Lori Garzio, on 1/14/19

New Note

Metadata Start Date End Date Comment
GS01SUMO-SBD12
12/13/15, 7:00 PM 12/11/16, 7:00 PM

Upon recovery of deployment 2, damage to some components of the surface buoy were noted (e.g., FDCHP was bent, the tops of the METBK-PRC and FleetBroadBand antenna were missing).

Id: 95 By: lgarzio

GS01SUMO
12/8/17, 7:00 PM 12/8/18, 11:42 PM

Deployment 3: *UPDATED 2020-04-28: This mooring was recovered 2018-12-09 which was a year later than originally intended. Some instruments were able to continue sampling for the extended deployment.* Recovery of the mooring was planned during the NBP 17-09 cruise (Nov - Dec 2017) but not completed due to adverse weather conditions. Recovery options are pending. Instruments that are still operational are likely reaching (or have surpassed) their deployment lifespan.

Id: 834 By: lgarzio

GS01SUMO-SBD12-02-FLORTD000
5/1/19, 8:00 PM 1/20/20, 5:01 AM

Deployment 4: There have been some gaps in the telemetered FLORD data beginning 2019-05-02. Root cause still to be determined.

Id: 1584 By: cdobson